2.3. CE Mark Compliance

SPECIFICATIONS, APPROVALS AND WARRANTY

M102E/M501 TRS

(Addendum to M101E Manual - P/N 04740 Rev A)

Sample filter: Equipped with PTFE filter element in the internal filter assembly.

Sample flow of 650 +/- 65 cc/min.

Vacuum pump (internal or external) capable of 14"Hg Absolute pressure @ 1 slpm or better.

Software settings:

Dynamic span

OFF

Dynamic zero

OFF

Dilution factor

OFF

AutoCal

ON or OFF

Dual range

ON or OFF

Auto-range

ON or OFF

Temp/Pressure compensation

ON

Under the designation, the analyzer may be operated with or without the following optional equipment:

Rack mount with chassis slides.

Rack mount without slides, ears only.

Zero/span valve options.

Internal zero/span (IZS) option with either:

SO2 permeation tube - 0.4 ppm at 0.7 liter per minute; certified/uncertified.

SO2 permeation tube - 0.8 ppm at 0.7 liter per minute; certified/uncertified. Under the designation, the IZS option cannot be used as the source of calibration.

4-20mA isolated analog outputs.

Status outputs.

Control inputs.

RS-232 output.

Ethernet output.

Zero air scrubber.

4-20mA, isolated output.

2.3. CE Mark Compliance

See Section 2.3 of the M101E Manual - P/N 04740 Rev A

User Notes:

4

05514 Rev A1

Page 10
Image 10
Teledyne manual CE Mark Compliance, User Notes, M102E/M501 TRS, Specifications, Approvals And Warranty, Rev A1