Table 10-4 – Temperature Controller – Primary Parameter Settings

M102E/M501 TRS

 

TROUBLESHOOTING & REPAIR

(Addendum to M101E Manual - P/N 04740 Rev A)

 

Table 10-4 – Temperature Controller – Primary Parameter Settings

DISPLAY

PARAMETER

SET TO

COMMENTS

NAME

 

 

 

 

 

 

Sets the controller to reverse action (hotter

 

OFF

0

temperatures = less heating; lower temperatures =

 

more heating) and turns off the heater should the

 

 

 

 

 

 

thermocouple input fail.

 

Input type

3

K thermocouple

 

Digital Filter

5

Sets the response time of the digital noise filter to 5

 

sec.

 

 

 

 

Lower input range

32

Sets the lower end of the controller full scale bandwidth

 

limit

to 32˚C

 

 

 

Upper input range

1100

Sets the upper end of the controller full scale

 

limit

bandwidth to 1100˚C

 

 

 

 

 

Sets the High alarm point to trigger when the process

 

High Alarm Type

10

value exceeds the high alarm set point only after the

 

process value rises above the lower limit for the first

 

 

 

 

 

 

time after power up

 

 

 

Sets the Low alarm point to trigger when the process

 

Low Alarm Type

10

value exceeds the high alarm set point only after the

 

process value rises above the lower limit for the first

 

 

 

 

 

 

time after power up

 

Alarm Hysteresis

3

Sets the hysteresis bandwidth for both alarm points to

 

3% of full scale.

 

 

 

 

Decimal Point

0

Sets the decimal point for PV & SV readings to NONE

 

Process variable

0

OFF

 

Offset

 

 

 

 

Set variable

0

OFF

 

Offset

 

 

 

 

Celsius/Fahrenheit

C

Sets controller to perform function on Celsius

 

temperature scale.

 

 

 

 

Fuzzy Logic

ON

Turns of the controllers advanced fuzz logic feature

 

which further reduces overshoot and fluctuations.

 

 

 

Ignore any other parameters that may appear.

10.7. Technical Assistance

If this addendum and its trouble-shooting / repair sections do not solve your problems, technical assistance may be obtained from Teledyne Instruments, Customer Service, 6565 Nancy Ridge Drive, San Diego, CA 92121. Phone: +1 858 657 9800 or 1-800 324 5190. Fax: +1 858 657 9816. Email: api-customerservice@teledyne.com.

Before you contact customer service, fill out the problem report form in Appendix C, which is also available online for electronic submission at http://www.teledyne-api.com/forms/index.asp.

05514 Rev A1

67

Page 73
Image 73
Teledyne manual Technical Assistance, M102E/M501 TRS, Troubleshooting & Repair, Rev A1