Option

GETTING STARTED

M102E/M501 TRS

 

(Addendum to M101E Manual - P/N 04740 Rev A)

MODEL 700

Gas Dilution Calibrator

(with O3 generator option)

VENT

 

MODEL 701

Calibrated

Zero Air

H2S gas

Generator

(At high

 

concentration)

 

 

Source of

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SAMPLE Gas

 

 

 

 

 

 

VENT if input is pressurized

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Sample

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Exhaust

 

 

 

 

 

 

 

 

MODEL

 

 

 

 

 

 

 

 

 

 

 

External Zero

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

102E

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Air Scrubber

 

Filter

 

Zero Air

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M501-TRS

To AnalyzerValve OptionManual background To Converter

From AnalyzerManual backgroundManual background From Converter

Figure 3-10: Pneumatic Connections for Formal Calibration of Units with an IZS Valve

Option

Source of

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SAMPLE Gas

 

 

 

 

 

VENT if input is pressurized

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Sample

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Exhaust

 

 

 

 

 

 

 

 

MODEL

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

102E

 

 

 

 

 

 

 

 

 

Span

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Ambient

 

 

 

 

 

 

 

 

 

 

Zero Air (Scrubber)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Air

 

 

 

 

 

 

 

 

 

 

 

M501-TRS

To AnalyzerManual backgroundManual background To Converter

From AnalyzerManual backgroundManual backgroundManual background From Converter

Figure 3-11: Pneumatic Connections for Informal Calibration Checks of Units with IZS

Valve Option

NOTE

Gas flow must be maintained at all times for units with IZS Options installed. The IZS option requires a permeation tube ( customer supplied ) which emits H2S. Insufficient gas flow can build up H2S to levels that will damage the instrument.

Remove the permeation device when taking the analyzer out of operation.

16

05514 Rev A1

Page 22
Image 22
Teledyne manual Model, M102E/M501 TRS, Getting Started, Valve Option, Rev A1