Teledyne manual M102E/M501 TRS, Troubleshooting & Repair, Rev A1

Models: 102E

1 90
Download 90 pages 13.69 Kb
Page 72
Image 72
Table 10-3 – Temperature Controller – Primary Parameter Settings

TROUBLESHOOTING & REPAIR

 

M102E/M501 TRS

 

 

 

(Addendum to M101E Manual - P/N 04740 Rev A)

 

Table 10-3 – Temperature Controller – Primary Parameter Settings

DISPLAY

PARAMETER

SET TO

COMMENTS

NAME

 

 

 

 

Proportional

11

Sets the bandwidth of the proportional control function

 

Band

to ± 11% of full scale

 

 

 

Integral time

10

Sets the reaction time for the proportional control

 

function to 10 seconds.

 

 

 

 

Derivative Time

7.7

Sets the reaction time for the derivative control

 

function (which reduces overshoot) to 7.7 seconds.

 

 

 

 

Low Alarm Limit

950

Sets the low alarm point to 950˚C

 

High Alarm Limit

1050

Sets the high alarm point to 1010˚C

 

 

 

Sets the cycle time (which is divided proportionally

 

 

 

between On and Off) for 2 seconds.

 

Cycle Time

2

EXAMPLE: for a duty cycle of 25% with a Cycle time of

 

 

 

2 seconds, the controller will turn the heater on for 0.5

 

 

 

seconds every 2 seconds

 

 

 

Sets the bandwidth of the area around the set point

 

Hysteresis

3

where the controller will not try to change the process

 

value (thereby inducing unnecessary fluctuations) to 3

 

 

 

 

 

 

% of full scale.

 

Autotune

0 (OFF)

Autotune is OFF

 

Can be either

 

 

 

0

All parameters are changeable

 

Parameter Lock

All parameters are locked

 

1

 

 

 

2The set value can be changed but all other parameters are locked

Ignore any other parameters that may appear.

66

05514 Rev A1

Page 72