Be Careful!

TROUBLESHOOTING & REPAIR

M102E/M501 TRS

 

(Addendum to M101E Manual - P/N 04740 Rev A)

2.Allow it to cool to room temperature. This may take up to 30 minutes.

3.Remove the instruments top cover (see Section 3.2 of this addendum)

4.Remove the four screws holding the converter cover in place.

5.Carefully lift the converter cover away.

6.Loosen front and rear pneumatic fittings at each end of the tube and remove the gas lines from the converter tube.

7.Remove the front ceramic bobbin from the converter tube. The thermistor assembly will come with it.

NOTE:

Be Careful!

The ceramic bobbins at each end of the heater assembly are fragile.

8.Slide a new tube into the heater coil assembly.

9.Make sure that the thermistor is threaded through its notch on the center hole of the front converter bobbin and properly seated in the corresponding indentation in the body of the quartz heater tube.

10.Reattach the front bobbin onto the converter tube.

11.Reattach the gas lines and retighten the pneumatic fittings.

12.Reassemble the converter cover and reattach the M501-TRS top cover.

NOTE

The M501-TRS will not operate properly with the top cover removed.

The air cooling required to stabilize the temperature of the converter tube is dependent

on air flow patterns that only exist with the top cover in place.

Without the top cover in place, the thermal cutout may overheat, and shut the heating

element off.

13.Restart the M501-TRS.

14.Check the converter efficiency. See Section 10.4.2 of this addendum.

10.6.Manually Programming the M501-TRS Temperature Controller

NOTE

The temperature controller has been programmed at the factory and should not be

altered, (except for temperature set point).