Teledyne Replacing the PMT, HVPS or TEC, M102E/M501 TRS, Troubleshooting & Repair, Rev A1

Models: 102E

1 90
Download 90 pages 13.69 Kb
Page 65
Image 65
10.5.3. Replacing the PMT, HVPS or TEC

M102E/M501 TRS

TROUBLESHOOTING & REPAIR

(Addendum to M101E Manual - P/N 04740 Rev A)

5.Carefully remove the UV filter.

6.Install the UV filter. Handle carefully and never touch the filter’s surface. The UV filter’s wider ( ring ) side should be facing out. Install the UV filter retainer and tighten screws.

7.Install the shutter cover and minifit connector. Tighten 4 screws.

8.Re-plug J4 connector into the motherboard.

10.5.3. Replacing the PMT, HVPS or TEC

PMT Housing End Plate

This is the entry to the PMT Exchange

PMT Output

 

Connector

PMT Preamp PCA

PMT Power Supply

High voltage Power Supply

& Aux. Signal

(HVPS)

Connector

 

 

PMT

 

O-Test LED

PMT Cold Block

Insulation Gasket

PMT Temperature

Light from Reaction

Chamber shines

Sensor

through hole in side

 

of Cold Block

 

Thermo-Electric Cooler

 

(TEC)

 

PMT Heat Exchange Fins

 

TEC Driver PCA

Cooling Fan

Housing

Connector to PMT

Pre Amp PCA

12V Power

Connector

Figure 10-3: PMT Assembly - Exploded View

The PMT should last for the lifetime of the analyzer. However, in some cases, the high voltage power supply (HVPS) or the thermo-electric cooler (TEC) may fail. To replace the PMT, the HVPS or the TEC:

1.Power down the analyzer, disconnect the power cord, remove the cover and disconnect all pneumatic and electrical connections from the sensor assembly.

2.Remove the entire sensor module assembly from the.

05514 Rev A1

59

Page 65
Image 65
Teledyne manual Replacing the PMT, HVPS or TEC, M102E/M501 TRS, Troubleshooting & Repair, 3 PMT Assembly - Exploded View