9.4.1.1. Sample Chamber

THEORY OF OPERATION

M102E/M501 TRS

 

(Addendum to M101E Manual - P/N 04740 Rev A)

9.4.1.1. Sample Chamber

The main electronic components of the sample chamber are the reference detector(see Section

10.2.2of the M101E Manual - P/N 04740 Rev A); the UV Lamp (see Section 10.2.1 of the M101E Manual - P/N 04740 Rev A) and its electronically operated shutter (see Section 9.2.1 of this addendum); and the sample chamber heating circuit,

UV Source Lamp

Shutter Housing

UV Source Lens &

Housing

PMT Lens &

Sample Air Housing

Outlet

O-Ring O-Ring

Seal Seal

Shutter Assy

 

Sample Air

(hidden from view)

 

Inlet

Sample Chamber

 

Reference

Heater

 

 

Detector

 

 

Sample Chamber

 

 

Sample Chamber

 

O-Ring

Temperature Sensor

 

 

Seal

 

 

 

Sample Chamber

Light Trap

 

Heater

 

 

Figure 9-3: M102E Sample Chamber

9.4.1.2. Sample Chamber Heating Circuit

In order to reduce temperature effects, the sample chamber is maintained at a constant 50°C, just above the high end of the instrument’s operation temperature range. Two AC heaters, one embedded into the top of the sample chamber, the other embedded directly below the reference detector’s light trap, provide the heat source. These heaters operate off of the instrument’s main AC power and are controlled by the CPU through a power relay on the relay board. A thermistor, also embedded in the bottom of the sample chamber, reports the cell’s temperature to the CPU through the thermistor interface circuitry of the motherboard.

48

05514 Rev A1

Page 54
Image 54
Teledyne Sample Chamber Heating Circuit, M102E/M501 TRS, Theory Of Operation, Sample Air, Inlet, Reference, O-Ring