3.6.1.2. M501-TRS Alarm Output Connections

M102E/M501 TRS

GETTING STARTED

(Addendum to M101E Manual - P/N 04740 Rev A)

TO change these settings, see Sections 6.9.4 and 6.7 of the M101E Manual - P/N 04740 Rev A respectively.

3.6.1.2. M501-TRS Alarm Output Connections

The rear panel of the M501-TRS includes a terminal strip by which connections can be made to the converters internal temperature alarm for more information on this alarm see Section 6.5).

Connect the input leads to your alarm-sensing device (e.g. datalogger) to the center two pins of the alarm output connector (see Figure 3-5).

Make sure the load does not exceed the rated capacity of the relay.

3.6.2. Pneumatic Connections:

This section supercedes the information contained in Section 3.1.2 of the M101E Manual - P/N 04740 Rev A.

CAUTION

To prevent dust from getting into the analyzer, it was shipped with small plugs inserted into each of the pneumatic fittings on the rear panel. Make sure that all dust plugs are removed before attaching exhaust and supply gas lines.

Sample and calibration gases should only come into contact with PTFE (Teflon) or glass materials. They should not come in contact with FEP or stainless steel materials.

Figures 3-7 and 3-8 show the most common configurations for gas supply and exhaust lines to the Model 102E Analyzer. Figures 3-9, 3-10 & 3-11 show the connections for units with valve options installed.

Please refer to Figures 3-1 & 3.3 for the location of pneumatic connections at the rear panel of the M102E and the M501-TRS.

Table 3-3: Inlet / Outlet Connector Nomenclature

 

 

 

M102E PNEMATIC CONNECTERS

 

REAR PANEL LABEL

 

FUNCTION

 

SAMPLE

 

Connects the sample gas to the analyzer. When operating the analyzer without

 

 

zero/span option, this is also the inlet for any calibration gases.

 

 

 

 

 

 

 

 

EXHAUST

 

Exhausts the gas sampled by the analyzer. Connect to an outside area away

 

 

from people.

 

 

 

 

 

 

 

 

SPAN

 

On units with zero/span/shutoff valve options installed, connect a gas line to

 

 

the source of calibrated span gas here.

 

 

 

 

 

 

 

 

ZERO AIR

 

On Units with zero/span valve or IZS option installed, this port connects the

 

 

zero air gas or the zero air cartridge to the analyzer.

 

 

 

 

 

 

 

 

TO CONVERTER

 

Sample gas leaves the M102E to be conditioned by the M501-TRS via this port.

 

 

 

 

 

FROM CONVERTER

 

Sample gas returns to the M102E after being conditioned by the M501-TRS via

 

 

this port.

 

 

 

 

 

 

 

 

 

 

M501-TRS PNEMATIC CONNECTERS

 

REAR PANEL LABEL

 

FUNCTION

 

 

 

 

05514 Rev A1

11

Page 17
Image 17
Teledyne Pneumatic Connections, 3.6.1.2. M501-TRS Alarm Output Connections, M102E PNEMATIC CONNECTERS, M102E/M501 TRS