INSTRUMENT MAINTENANCE

 

M102E/M501 TRS

 

 

 

 

 

 

INSTRUMENT MAINTENANCE

 

(Addendum to M101E Manual - P/N 04740 Rev A)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

8. INSTRUMENT MAINTENANCE

 

 

 

 

 

 

 

 

The following table supercedes Table 9-1 of the M101E Manual - P/N 04740 Rev A

 

 

 

 

 

 

 

 

 

 

Table 8-1:

M102E Preventive Maintenance Schedule

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ITEM

ACTION

FREQUENCY

CAL

M101E

 

 

DATE PERFORMED

 

 

 

 

CHECK

MANUAL

 

 

 

 

 

 

 

 

 

 

 

 

 

SECTION

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M501 SO2

Replace scrubber

 

 

8.1.1.3 of

 

 

 

 

 

 

 

 

 

As required

Yes

this

 

 

 

 

 

 

 

 

 

scrubber

material

 

 

 

 

 

 

 

 

 

 

 

addendum

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1Particulate filter

Change particle

Weekly

No

9.3.1

 

 

 

 

 

 

 

 

 

 

filter

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Verify test

Review and

Weekly

No

Appendix C

 

 

 

 

 

 

 

 

 

functions

evaluate

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Zero/span check

Evaluate offset

Weekly

--

7.3, 7.6, 7.9

 

 

 

 

 

 

 

 

 

and slope

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1Zero/span

Zero and span

Every 3 months

--

7.2, 7.4, 7.5,

 

 

 

 

 

 

 

 

 

calibration

calibration

7.7, 7,8

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1External zero air

Exchange

Every 3 months

No

9.3.4

 

 

 

 

 

 

 

 

 

scrubber (option)

chemical

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1Perform flow

Check Flow

Every 6 Months

No

11.5.2

 

 

 

 

 

 

 

 

 

check

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1Sample

Clean windows

Annually or as

Yes

9.3.6

 

 

 

 

 

 

 

 

 

chamber optics

and filters

necessary

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1Critical flow

 

 

 

 

 

 

 

 

 

 

 

 

 

orifice & sintered

Replace

Annually

Yes

9.3.7

 

 

 

 

 

 

 

 

 

filters

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Internal IZS

Replace

Annually

YES

9.3.2

 

 

 

 

 

 

 

 

 

Permeation Tube

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Perform

 

Annually or after

 

 

 

 

 

 

 

 

 

 

 

pneumatic leak

Verify Leak Tight

repairs involving

Yes

11.5.1

 

 

 

 

 

 

 

 

 

check

 

pneumatics

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2Pump

 

 

 

See

 

 

 

 

 

 

 

 

 

Replace

Every 2 years, or

Yes

instruction in

 

 

 

 

 

 

 

 

 

diaphragm

as necessary

diaphragm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

kit

 

 

 

 

 

 

 

 

 

PMT sensor

Low-level

On PMT/ preamp

 

 

 

 

 

 

 

 

 

 

 

changes if

 

 

 

 

 

 

 

 

 

 

 

hardware

hardware

Yes

11.6.3

 

 

 

 

 

 

 

 

 

0.7 < SLOPE or

 

 

 

 

 

 

 

 

 

calibration

calibration

 

 

 

 

 

 

 

 

 

 

 

SLOPE >1.3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1 These Items are required to maintain full warranty, all other items are strongly recommended.

2 A pump rebuild kit is available from Teledyne Instruments Customer Service including all instructions and required parts (see Appendix B for part numbers).

05514 Rev A1