The M501-TRS controller includes an auto tune feature which allows the controller to find and set optimum values for various process control parameters so that the controller can establish and maintain the converter oven at the temperature set value in the most stable and efficient manner.

NOTE

Before initiating the autotune feature make sure that the converter temperature oven

has reached a stable, constant temperature.

6.4.1. Initiating the Autotune Process:

1.Press the SELECT key once. The main display will show has reached a stable, constant temperature.6.4.1. Initiating the Autotune Process:

2.Use the SELECT , DOWN or 100’s UP key to scroll through the primary menu parameters until

the display shows Manual backgroundManual backgroundManual background (AT =Autotune).

3.Press the DATA key once. The display will show Manual backgroundManual background (zero = Off).

4.Press the 1’s UP key once. The display will show Manual background (1 = autotune based on set point value).

5.Press the ENTER key to begin the autotune process. A blinking decimal point will appear at the bottom right-hand corner of the main display.

6.Wait until the blinking light stops. This may take up to 30 minutes.

7.The autotune process is finished. The autotune parameter value will automatically reset to zero (off).

8.Press the PV/SV mode key to return to operational mode.

Note

The P-I-D parameters calculated by autotuning will be retained even if the power is lost. However, if the power is turned off during the auto-tuning process, you must restart autotuning.