APPENDIX A-3: Warnings and Test Functions, Revision A.2

M102E/M501 TRS

APPENDIX A-3: Warnings and Test Functions, Revision A.2

(Addendum to M101E Manual - P/N 04740 Rev A)

 

 

 

 

 

APPENDIX A-3: Warnings and Test Functions, Revision A.2

NOTE

Warning messages and test functions for the M102E are the same as those for the M102E (see Appendix A-3 of the M101E Manual - P/N 04740110 Rev A) with the following exceptions:

Table A-4: Warning Messages deleted from M102 Software Revision A.2

NAME

MESSAGE TEXT

DESCRIPTION

 

 

 

WCONVTEMP

CONV TEMP WARNING

Converter temperature outside of warning limits specified

 

 

by CONV_SET variable.

 

 

 

Table A-5: Test Functions Changed and Added to M102 Software Revision A.2

TEST Function

Message Text

DESCRIPTION

 

 

 

STABILITY

TRS STB1=0.0 PPB 1

Concentration stability (standard deviation based on

 

 

setting of STABIL_FREQ and STABIL_SAMPLES).

 

 

 

SO2 SLOPE

SO2 SLOPE2=1.061

Slope for current SO2 measurement range, computed

 

 

during zero/span calibration.

 

 

 

SO2 OFFSET

SO2 OFFS2=250.0 MV

Offset for current TRS measurement range, computed

 

 

during zero/span calibration.

 

 

 

TRS SLOPE

TRS SLOPE2=1.061

Slope for current SO2 measurement range, computed

 

 

during zero/span calibration.

 

 

 

TRS OFFSET

TRS OFFS2=250.0 MV

Offset for current TRS measurement range, computed

 

 

during zero/span calibration.

 

 

 

1 Shown as it appear when analyzer is in TRS mode. In SO2 mode appear as SO2 STB. In multigas mode, both versions appear.

2 Only Appears in SO2 and multigas measurement modes.

2 Only Appears in TRS and multigas measurement mode.

Table A-6: Test Functions Deleted from M102 Software Revision A.2

TEST Function

Message Text

DESCRIPTION

 

 

 

RESPONSE 2

RSP=1.11(0.00) SEC

Instrument response. Length of each signal processing

 

 

loop. Time in parenthesis is standard deviation.

 

 

 

VACUUM 5

VAC=9.1 IN-HG-A

Vacuum pressure.

RCELLDUTY

RCELL ON=0.00 SEC

Sample chamber temperature control duty cycle.

 

 

 

IZSDUTY

IZS ON=0.00 SEC

IZS temperature control duty cycle.

 

 

 

CONVTEMP

CONV TEMP=315.0 C

TRS Æ SO2 Converter temperature.

05515 Rev A.2

A-7