5.4.2.2. Setting Hessen Protocol Status Flags

M102E OPERATING INSTRUCTIONS

M102E/M501 TRS

 

(Addendum to M101E Manual - P/N 04740 Rev A)

5.4.2.2. Setting Hessen Protocol Status Flags

The following table supercedes Table 6-29 of the M101E Manual - P/N 04740 Rev A/

Table 5-6: Default Hessen Status Bit Assignments

STATUS FLAG NAME

DEFAULT BIT

 

ASSIGNMENT

 

 

WARNING FLAGS

 

 

 

SAMPLE FLOW WARNING

0001

 

 

PMT DET WARNING

0002

 

 

UV LAMP WARNING

0002

 

 

HVPS WARNING

0004

 

 

DARK CAL WARNING

0008

 

 

RCELL TEMP WARNING

0010

 

 

IZS TEMP WARNING

0020

 

 

PMT TEMP WARNING

0040

 

 

CONV TEMP WARNING

1000

 

 

OPERATIONAL FLAGS

 

 

 

In Manual Calibration Mode

0200

 

 

In Zero Calibration Mode

0400

 

 

In Span Calibration Mode

0800

 

 

UNITS OF MEASURE FLAGS

 

 

 

UGM

0000

 

 

MGM

2000

 

 

PPB

4000

 

 

PPM

6000

 

 

SPARE/UNUSED BITS

0080, 0100, 1000,

 

8000

 

 

UNASSIGNED FLAGS

Box Temp Warning

Front Panel Warning

 

 

Sample Press Warning

Analog Cal Warning

 

 

System Reset

Cannot Dyn Zero

 

 

Rear Board Not Detected

Cannot Dyn Span

 

 

Relay Board Warning

Invalid Conc

 

 

User Notes: