5.4. Remote Operation of the Analyzer

M102E OPERATING INSTRUCTIONS

M102E/M501 TRS

(Addendum to M101E Manual - P/N 04740 Rev A)

5.4. Remote Operation of the Analyzer

5.4.1. Control Inputs

The description of the control inputs in Section 6.12.1.2 of the M101E Manual - P/N 04740 Rev A in correctly shows an external low span cal input. This is not correct. Neither the M102E nor the M102E has an external low span cal input.

The following table and figures supercede Table 6-23 and Figures 6-18 & 6-19 of the M101E Manual - P/N 04740 Rev A respectively.

 

 

Table 5-4:

M102E Control Input Pin Assignments

 

 

 

 

INPUT

STATUS

CONDITION WHEN ENABLED

 

 

 

 

A

External Zero Cal

Zero calibration mode is activated. The mode field of the

display will read ZERO CAL R.

 

 

 

 

 

 

 

B

External Span Cal

Span calibration mode is activated. The mode field of the

display will read SPAN CAL R.

 

 

 

C, D, E, & F

 

Unused

 

 

 

 

 

 

Digital Ground

Provided to ground an external device (e.g., recorder).

 

 

 

 

 

 

 

 

U

DC Power For

Input for +5 VDC required to activate inputs A - F. This voltage

Input Pull Ups

can be taken from an external source or from the “+” pin.

 

 

+Internal +5v Internal source of +5V which can be used to activate inputs

Supply when connected to pin U.

CONTROL IN

A

B

C

D

E

F

U

+

ZERO

SPAN5.4.1. Control Inputs

Figure 5-2: Control Inputs with local 5 V power supply

28

05514 Rev A1

Page 34
Image 34
Teledyne Remote Operation of the Analyzer, Control Inputs, Zero, M102E/M501 TRS, M102E OPERATING INSTRUCTIONS, Rev A1