Teledyne 102E manual APPENDIX A-2 Setup Variables For Serial I/O, Revision A.2, Table A-2, Rev A.2

Models: 102E

1 90
Download 90 pages 13.69 Kb
Page 80
Image 80
APPENDIX A-2: Setup Variables For Serial I/O, Revision A.2

APPENDIX A-2: Setup Variables For Serial I/O, Revision A.2M102E/M501 TRS

(Addendum to M101E Manual - P/N 04740 Rev A)

APPENDIX A-2: Setup Variables For Serial I/O, Revision A.2

NOTE

Setup Variables for the M102E are the same as those for the M102E (see Appendix A-3

of the M101E Manual - P/N 04740110 Rev A) with the following exceptions:

Table A-1: Changed or added Setup Variables for M102E Software Revision A.2

 

SETUP VARIABLE

 

NUMERIC

DEFAULT

 

VALUE RANGE

 

 

DESCRIPTION

 

 

 

 

 

 

 

 

 

 

UNITS

VALUE

 

 

 

 

 

 

 

MEASURE_MODE

SO2-TRS

 

SO2, SO2-TRS,

 

Gas measurement mode. Enclose

 

 

 

 

 

 

TRS

 

value in double quotes (") when

 

 

 

 

 

 

 

setting from the RS-232 interface.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TRS_SPAN1

Conc

400

 

0.1–50000

 

Target TRS concentration during

 

 

 

 

 

 

 

 

 

span calibration of range 1.

 

 

 

 

 

 

 

 

 

TRS_SPAN2

Conc

400

 

0.1–50000

 

Target TRS concentration during

 

 

 

 

 

 

 

 

 

span calibration of range 2.

 

 

 

 

 

 

 

 

 

TRS_SLOPE1

PPB/mV

1

 

0.25–4

 

TRS slope for range 1.

 

 

 

 

 

 

 

 

 

TRS_SLOPE2

PPB/mV

1

 

0.25–4

 

TRS slope for range 2.

 

 

 

 

 

 

 

 

 

TRS_OFFSET1

mV

0

 

-1500–1500

 

TRS offset for range 1.

 

 

 

 

 

 

 

 

 

TRS_OFFSET2

mV

0

 

-1500–1500

 

TRS offset for range 2.

 

 

 

 

 

 

 

 

 

CE_FACTOR1

1

 

0.8–1.2

 

Converter efficiency factor for TRS

 

 

 

 

 

 

 

 

 

for range 1.

 

 

 

 

 

 

 

 

 

CE_FACTOR2

1

 

0.8–1.2

 

Converter efficiency factor for TRS

 

 

 

 

 

 

 

 

 

for range 2.

 

 

 

 

 

 

 

 

 

REMOTE_CAL_MODE

SO2-

 

SO2-LOW,

 

Gas and range to calibrate during

 

 

 

 

LOW

 

SO2-HIGH,

 

contact-closure and Hessen

 

 

 

 

 

 

 

calibration. Enclose value in double

 

 

 

 

 

 

TRS-LOW,

 

 

 

 

 

 

 

 

quotes (“) when setting from the RS-

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TRS-HIGH

 

232 interface.

 

 

 

 

 

 

 

 

 

STABIL_GAS

SO2

 

SO2,

 

Gas to use to measure concentration

 

 

 

 

 

 

TRS

 

stability. Enclose value in double

 

 

 

 

 

 

 

quotes (") when setting from the RS-

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

232 interface.

 

 

 

 

 

 

 

 

 

 

 

 

Table A-2:

Deleted Setup Variables for M102E Software Revision A.2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP VARIABLE

 

 

NUMERIC

 

 

DEFAULT VALUE

 

 

VALUE

 

 

DESCRIPTION

 

 

 

 

 

UNITS

 

 

 

 

 

RANGE

 

 

 

 

 

CONV_TYPE 2

 

 

MOLY

 

NONE,

 

Converter type.

 

 

 

 

 

 

 

 

 

 

MOLY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CONV_SET 2

 

ºC

315

 

 

0–350

 

Converter temperature set point and

 

 

 

 

 

 

 

Warnings:310–320

 

 

 

 

warning limits.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TEST_CHAN_ID

 

 

NONE

 

CONV

 

All other Test Channel settings are

 

 

 

 

 

 

 

 

 

 

TEMP ,

 

the same as those listed in appendix

 

 

 

 

 

 

 

 

 

 

 

 

 

A-1 of the M101E Manual - P/N

 

 

 

 

 

 

 

 

 

 

 

 

 

04740110 Rev A

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A-6

05515 Rev A.2

Page 80
Image 80
Teledyne 102E manual APPENDIX A-2 Setup Variables For Serial I/O, Revision A.2, Table A-2, Rev A.2, Measuremode, Stabilgas