6.3.To Manually Adjust the Converter Oven Temperature:

M102E/M501 TRS

M501-TRS OPERATING INSTRUCTIONS

(Addendum to M101E Manual - P/N 04740 Rev A)

6.3.To Manually Adjust the Converter Oven Temperature:

CAUTION !

DO NOT SET THE TEMPERATURE HIGHER THAN 1050OC

1.Set the main display to show the current value of the set variable by pressing the PV/SV mode key.

2.To set each digit:

a.Press the up-arrow under that digit once. The digit will flash.

b.to increment that digit, press and hold the digit until the appropriate number is displayed.

c.To decrement that digit press and hold the DOWN key until the appropriate number is displayed.

d.To increment/decrement the 1000’s digit it is necessary to adjust increment/decrement the 100’s digit up and down. Each time the 100’s digit passes “0” the 1000s digit will increment or decrement correspondingly.

3.Once the desired value is reached, press the ENT key to store the new set value

4.Return the main display to process mode by pressing the PV/SV mode key once

EXAMPLE to change the set value from 950 to equal 1010.

ACTION

RESULT

 

 

Press the PV/SV mode key

The SV indicator will lit up and the display will

 

show 950.

 

 

Press the 10’s UP key once

The 10’s digit will begin to blink

 

 

Press the DOWN key

The 10’s digit will decrement from “5”. Release

 

the DOWN key when the 10’s digit reads “1”.

 

 

Press the 100’s UP Key once

The 100’s digit will begin to blink

 

 

Press and hold the 100’s UP key

The 100’s digit will increment from “9”. When it

 

passes “0” the 1000’s digit will increment to “1”.

 

Release the 100’s UP key.

 

 

--

The Display should now read “1010”

 

 

Press the ENTER key

The new set value is recorded

 

 

Press the PV/SV mode key

The current level of the process value will be

 

displayed.

 

 