Teledyne Specifications, Approvals And Warranty, EPA Equivalency Designation, M102E/M501 TRS

Models: 102E

1 90
Download 90 pages 13.69 Kb
Page 9
Image 9
SPECIFICATIONS, APPROVALS AND WARRANTY

M102E/M501 TRS

SPECIFICATIONS, APPROVALS AND WARRANTY

(Addendum to M101E Manual - P/N 04740 Rev A)

2.SPECIFICATIONS, APPROVALS AND WARRANTY

2.1. Specifications

There are no significant differences between the performance specifications for the M102E and the M102E as listed in Section 2.1 of the M101E Manual - P/N 04740 Rev A.

2.1.1. M501-TRS Specifications

Table 2-1: Model 102E Basic Unit Specifications

Minimum Converter Efficiency

H2S >95%

 

COS >90%

 

 

CS2 >90%

 

Maximum TS Concentration for

20 ppmv

 

specified conversion efficiency

 

 

 

 

 

Sample Flow Rate

650cc/min. ±10% - driven by M102E pneumatic system

 

 

 

Optimum Converter

1000°C (factory setup)

Temperature

 

 

 

 

 

Maximum Converter

1100°C

 

Temperature

 

 

 

 

 

Dimensions H x W x D

7" x 17" x 22" (178 mm x 432 mm x 559 mm)

 

 

 

Weight

16 lbs (8 kg)

 

 

26 lbs (12 kg) CE version

 

 

 

AC Power Rating

115 V, 50/60 Hz

- 400 Watts;

230 V, 50/60 Hz

- 575 Watts; CE Version

 

 

 

Internal Alarms

High Alarm Point: 1050°C

Low Alarm Point:

950°C

 

 

 

Alarm Output Relay

SPST - 1 point: Alarm output is energized should either the temperature

controller’s high or low internal alarm set points be activated.

 

 

 

Alarm Output Rating

220V AC/30V DC, 1A (resistive load)

 

 

Environmental

Installation category (over-voltage category) II; Pollution degree 2

 

 

Certifications

IEC 1010-1 / 61010-1:93 (includes A1) + A2:95,

 

 

 

For indoor use at altitudes ≤ 2000m only

2.2. EPA Equivalency Designation

No EPA equivalency standards exist for TRS measurement, however, the M102E analyzer qualifies for EPA equivalency designation as Reference Method Number EQSA-0495-100 per 40 CFR Part 53 when operated under the following conditions:

Measurement Mode: SO2 single gas mode.

Range: Any range from 50 parts per billion (ppb) to 10 parts per million (ppm).

Ambient temperature range of 5 oC to 40 oC.

Line voltage range of 105-125 VAC or 220-240 VAC, at 50 or 60 Hz.

05514 Rev A1

3

Page 9
Image 9
Teledyne 102E Specifications, Approvals And Warranty, EPA Equivalency Designation, 2.1.1. M501-TRS Specifications, Rev A1