NDIR Gas Analyzer Specifications

Teledyne Analytical Instruments 89
9. Specifications

9.1 General Specifications

9.1.1 Standard Specifications

Measurement Method:
NO, SO2, CO2, CO, CH4:
Non-dispersion infrared-ray absorption
method
Single light source and single beams (single
beam system)
O2:
Fuel cell O2 analyzer (built-in) or
paramagnetic O2 analyzer (built-in) or
zirconia O2 analyzer (externally installed
TYPE: ZFK7)
Measured Gas/Range:
Min Range Max Range
NO 0-200 ppm 0-5000ppm
SO2 0-200 ppm 0-10 vol.%
CO2 0-100 ppm 0-100 vol.%
CO 0-200 ppm 0-100 vol.%
CH4 0-500 ppm 0-100 vol.%
O2
(built in)
Fuel Cell
0-10 vol.% 0-25 vol.%
O2
(built in)
Paramag.
0-5 vol.% 0-100 vol.%
None 100-95 vol.%