NDIR Gas Analyzer | Specifications | |
|
|
|
Allowable load 100KΩ for 0 to 1V DC
∙Refer to Table2 for the channel No. of displayed values and analog output signals.
Analog Input Signal:
For signal input from externally installed
O2 sensor.
Signal requirement:
(1)Signal from Zirconia O2 sensor (TYPE: ZFK7)
(2)0 to 1V DC from an O2 sensor. Input section is not isolated. This feature is present when an O2 sensor is not built in.
*Externally installed O2 sensor must be purchased separately.
Digital Output: (option)
1C contact (24V DC/1A, resistive load) max.15 outputs
Instrument error, calibration error, range identification, auto calibration status, solenoid valve drive for auto calibration, High/Low limit alarm contact output.
All relay contacts are isolated mutually and from the internal circuit.
Digital Input: (option)
Voltage contact (supply
Remote range change over, auto calibration remote start, remote hold, average value reset, Isolated from the internal circuit with photocoupler.