NDIR Gas Analyzer

Settings and Calibration

 

 

 

confirm the execution of auto calibration. Press the ENT key to execute the auto calibration or press the ESC key to cancel.

6.4.2.2FORCED STOP OF AUTO CALIBRATION

This mode is used to stop the auto calibration forcibly.

1.In the “Set Auto Cal” screens point the cursor to “Auto Calibration Stop” using the “UP or DOWN key. Then press ENT. (“Auto Calibration Stop” appears when this screen is selected while auto calibration is performed.)

2.“Stop” is highlighted, displaying a message to confirm the stop of auto calibration. Press the ENT key to stop the auto calibration or press the ESC key to cancel the stop auto cal action.

“Auto Calibration” screen

Example

When Ch 1 and Ch2 are enabled for auto calibration (see Section 6.2.4):

Zero calibration

A message, “Zero cal.” blinks at Ch1 and Ch2.

Ch1 span calibration

A message, “Span cal.” blinks at Ch1.

Ch2 span calibration

A message, “Span cal.” blinks at Ch2.

Teledyne Analytical Instruments

53

Page 69
Image 69
Teledyne 7500ZA operating instructions Auto Calibration screen, Example