Teledyne 7500ZA operating instructions Example, SO2

Models: 7500ZA

1 126
Download 126 pages 9.92 Kb
Page 57
Image 57

NDIR Gas Analyzer

Settings and Calibration

 

 

 

2.Select the Ch you want to change by pressing the UP or DOWN key, then press the ENT key. The setting content is highlighted.

3.Toggle between “both” or “current” using the UP or DOWN key.

If “both” is selected, zero or span calibration will be performed on both Range 1 and Range 2 on the selected Ch when the calibration is performed.

If “current” is selected, zero or span calibration will be performed only for the range displayed when calibration is performed.

4.Press the ENT key after the selection, and the specified calibration is performed.

To close the Calibration Range Settings screen or to abort the entry midway, press the ESC key. The previous screen will return.

Example:

Ch1

Range 1: 0 to 200 ppm

both

NOx

Range 2: 0 to 2000 ppm

 

 

 

 

Ch 2

Range 1: 0 to 200 ppm

current

SO2

Range 2: 0 to 2000 ppm

 

Ch1: Range 1 and Range 2 are calibrated together

Ch2: Only the currently displayed range is calibrated

Note: When “both” is selected, the same calibration gas concentration must be used for both ranges.

Teledyne Analytical Instruments

41