Settings and Calibration

Model 7500ZA

 

 

 

6.5.2.1EXECUTION OF AUTO ZERO CALIBRATION (ONLY ONCE)

1.In the “Setting of Auto Zero Calibration” screen that appears, point the cursor to “Run” by pressing the UP or DOWN key. Then press ENT.

2.“Run” is highlighted, displaying a message to confirm execution of auto zero calibration. Press the ENT key to execute the calibration, or the ESC key to cancel.

6.5.2.2FORCED STOP OF AUTO ZERO CALIBRATION

This mode is used to forcibly cancel an auto zero calibration event.

1.In the “Setting of Auto Zero Calibration” screen that appears, point the cursor to “Stop” by pressing the UP or DOWN key, then press ENT. (“Auto Zero Calibration Stop” appears when the screen is selected while auto zero calibration is performed.)

2.“Stop” is highlighted, displaying a message to confirm the stop of auto zero calibration.

Press the ENT key to stop the auto zero calibration and the ESC key to cancel (not stopped).