NDIR Gas Analyzer | Settings and Calibration | |
|
|
|
Releasing peak count alarm
To release the peak count alarm, set the peak alarm to OFF. Turning on the peak alarm initiates the count at 0.
6.4 Setting Up the Auto Calibration
6.4.1 Auto Calibration
Auto calibration is automatically carried out at the time when zero span calibration are set. Before changing
the setting of auto calibration, set the auto calibration function to OFF.
1.Enter the “Setting of Auto Calibration” screen from the user mode, and the display shown at right appears. Use the UP and DOWN keys to align the cursor with the desired item and then press the ENT key.
2.In the “Set Auto Cal” screen that appears, enter the values for the various settings. To make a value entry or change, use the UP and DOWN keys to increment or decrement the value on digit at a time. Use the SIDE key to move the cursor to the next digit.
When all the settings are correct, press the ENT key to save them. Subsequent auto calibrations will use the saved settings.
Description of Auto Cal Settings
•Start Time : Setting for the first calibration event (day of the week, hour, minute)