NDIR Gas Analyzer Settings and Calibration
Teledyne Analytical Instruments 51
Auto calibration status contact output is closed during auto
calibration (NO side), and is open in other cases.
Example:
Start Time Sun 12:00
Cycle 1 day
Flow Time Zero 350 sec
CH 1 Span 350 sec
CH 2 Span 350 sec
CH 3 Span 350 sec
CH 4 Span 300 sec
CH 5 Span 300 sec
Ex. Time 300 sec
ON/OFF ON
Thus, with the above settings:
Setting Range:
Cycle: 1-99 hours or 1-40 days (default value—7days)
Flow Time: 60-90 sec (default value—300 sec)