NDIR Gas Analyzer

Settings and Calibration

 

 

 

Alarm Setting Descriptions

The alarm contact number can be assigned to any channel. Channels can also have multiple alarms assigned to them.

Channel:

Channel setting targeted for alarm.

H-Limit value:

Sets the high limit value (concentration) for

 

alarm.

 

L-Limit value:

Sets the low limit value (concentration) for alarm.

Kind of Alarm:

Select either High limit alarm, Low limit alarm,

 

and High limit or Low limit alarm, HH limit

 

alarm, and LL limit alarm.

 

High, HH:

Alarm contact closes when above

 

 

H-limit alarm.

 

Low, LL:

Alarm contact closes when below

 

 

L-limit alarm.

 

High or Low: Alarm contact closes when above

 

 

H-limit value or below lower limit

 

 

value.

ON/OFF:

Enables the alarm function when set to ON, or

 

disables it if set to OFF.

Note: The H-limit value cannot be set below the L-limit value, and the L-limit value cannot be set above the H-limit value.

If it is desired to set the H-limit value below the L-limit value already stored in the memory, reduce the L-limit value beforehand, and vice versa for setting the low limiot above the high limit.

Note: After turning on power, the alarm logic trigger is inactive for 10 minutes.

When an H-limit alarm occurs, the “H-alarm” message displays in the relevant component field. Other alarm displays are:

L-alarm for Low-limit alarm, HH- alarm for high limit alarm, HH-alarm for high-high alarm, and LL-alarm for low- low alarm.