NDIR Gas Analyzer | Settings and Calibration | |
|
|
|
Alarm Setting Descriptions
The alarm contact number can be assigned to any channel. Channels can also have multiple alarms assigned to them.
Channel: | Channel setting targeted for alarm. | |
Sets the high limit value (concentration) for | ||
| alarm. |
|
Sets the low limit value (concentration) for alarm. | ||
Kind of Alarm: | Select either High limit alarm, Low limit alarm, | |
| and High limit or Low limit alarm, HH limit | |
| alarm, and LL limit alarm. | |
| High, HH: | Alarm contact closes when above |
|
| |
| Low, LL: | Alarm contact closes when below |
|
| |
| High or Low: Alarm contact closes when above | |
|
| |
|
| value. |
ON/OFF: | Enables the alarm function when set to ON, or | |
| disables it if set to OFF. |
Note: The
If it is desired to set the
Note: After turning on power, the alarm logic trigger is inactive for 10 minutes.
When an
Teledyne Analytical Instruments | 45 |