Settings and Calibration | Model 7500ZA | |
|
|
|
Caution: When an auto calibration starts, the measurement screen appears automatically.
Any operation other than “Stop Auto Calibration” (see Item 6.4.2) is not permitted during auto calibration. “Stop Auto Calibration” cannot be performed with the key lock to ON.
To force a cancel of the auto calibration, set the key lock to OFF and then execute “Stop Auto Calibration”.
Turn power OFF and then ON again at the time set as the next start time in auto calibration, and then repeat it in the set cycle.
Remote Start
Whether the auto calibration is set at ON or OFF, an auto calibration is always available via a remote start input.
6.4.2 Forced Run/Stop Of Auto Calibration
Auto calibration can be performed just once or forcibly stopped while the calibration is in process
6.4.2.1EXECUTION OF AUTO CALIBRATION (ONLY ONCE)
1.In the “Set Auto Cal screen, move the cursor to “Auto Cal Run” using the UP or DOWN key, then press ENT.
2.“Run” is highlighted, displaying a message to
Teledyne Analytical Instruments | 52 |