Maintenance | Model 7500ZA | |
|
|
|
7.4.1.3REMOVING THE MEASURING UNIT
Refer to Fig.
5.Disconnect and remove detector output cables from detector output circuit board (No.12). Applying identification mark on top of removed cable connector will ensure proper pin assignment later.
6.Disconnect wiring to the
7.Detach the six screws (No. 16) fastening the base plate (No. 3) to remove the measuring unit. .
8.For assembly, reverse the disassembly procedures.
Note: Special care should be taken when assembling or disassembling the measuring cell to avoid using excessive force to the detector pipe or light source unit pipe. If the pipe is deformed or damaged by excessive force, there is a potential for gas leaks and this, in addition to creating a potential toxic hazard, will result in inaccurate measurement.
7.4.2Cleaning the Sample Cell
1.To clean the sample cell or IR transmission window, first clear away any dirt using a soft brush and then lightly wipe using a soft cloth. Do not use abrasive or paper towels.
Note: The window is fragile. Handle with care. Avoid scratching the window. Do not rub the dirt from the window roughly.
2.If the window or the sample cell interior is very dirty, use a soft
3.If the window is corroded, rub off any scale from the window lightly with a soft cloth to which chrome oxide powder is applied. If it is excessively corroded, it should be replaced.
4.When the cleaning operation is complete, assemble according to the sample cell assembly/disassembly procedures. Assemble the pipe carefully. If it becomes bent or damaged, replace it with a new part.
5.Do not wash the sample cell components with water.
Teledyne Analytical Instruments | 82 |