Settings and Calibration | Model 7500ZA | |
|
|
|
Station No.
This function allows the user to set the station No. for MODBUS communication. It is settable in the range from 00 to 32.
Range Setting
This function brings up a screen where the user can set the range for which concentration measurements are made. There are two settable ranges for each component each with a minimum and maximum value as requested at the time of purchase. The range to be used can be selected as Range 1 or Range 2.
To set the range:
1.Move the cursor to the item to be set using the UP or DOWN key, and then press ENT.
2.Move the cursor to the Ch (component) whose setting is to be changed using the UP or DOWN key, and then press ENT.
3.Move the cursor to the item whose setting is to be changed using the UP or DOWN key, and then press ENT.
The value for range 1 and range 2 must fall within the MIN and the MAX range of the instrument (including the MIN and the MAX range), and range 1 must be less than range 2.
4.Press the UP or DOWN key to change the value. Use the SIDE key to move to the next digit to the right. Press the ENT to save the new value. The measurement unit cannot be changed.
5.The decimal point can also be changed when it is highlighted. Use the UP or DOWN key to change the placement of the decimal point.
6.After all changes have been entered, press the ENT key to save.