Settings and Calibration | Model 7500ZA | |
|
|
|
6.2.4 Setting the Component/Range for Auto Calibration
In this section you can select the Ch (component) and the range to calibrate on for those components that have the auto ranging (AR) feature enabled.
1.From the User Mode screen select < Calibration parameters > → < Auto calibration component/range >. The “Auto Calibration Component Range” screen appears as shown.
2.Select the Ch you want to change by pressing the or the key. Press the ENT key. The cursor will be highlighted in the selected channel.
3.Select the range you want to use as the main calibration range by pressing the UP or DOWN key to move to the desired range.
4.Press the ENT key, and subsequent calibrations will be performed on the selected range when either auto calibration or auto zero calibration is requested.
Note regarding Auto Calibration
Component/Range Selection
The auto calibration or manual calibration of the component for which auto ranging (AR) is in effect will be performed on the range that is selected here. In this case, once the calibration is started, the range is automatically switched, and on completion of the calibration, the original range is resumed.