Specifications | Model 7500ZA | |
|
|
|
Zero gas ; Dry air or atmospheric air (Do not use with CO2 measurement)
Span gas ; For other than O2 measurement, each sample gas having concentration 90 to 100% of its measuring range.
For O2 measurement, O2 gas of 1 to 2 vol% /remainder N2 gas
2) Reverse range O2 measurement
Zero gas; 100vol% O2
Span gas ; For O2 measurement, O2 gas of 95.0%~95.5vol%/ remainder N2 gas
*For reverse range O2 measurement, infrared measurable component is not measurable.
9.1.7 Installation Requirements
•Indoor use (Select a place where the equipment does not receive direct sunlight, draft/rain or radiation from hot substances. If such a place cannot be found, a roof or cover should be prepared for protection).
•Avoid sites where unit receives heavy vibration.
•Select a site where atmospheric air is clean