NDIR Gas Analyzer

Settings and Calibration

 

 

 

Settings and Calibration

6.1 Range Switch

6.1.1Setting the Range Switch Mode

The range switch has three modes:

Manual (MR)

Remote (RR)

Auto Ranging (AR)

To set the range switch mode:

1.Press the MODE key while in measurement mode to display the User mode screen.

2.Move the cursor to “Switch Ranges” and press the ENT key.

3.In the “Channel Selection” screen that appears, move the cursor by pressing the UP or DOWN key, and select Ch (component).

4.Then press the ENT key.

5.The selected range switch mode will be highlighted. Press the UP or DOWN key to select the desired mode.

6.Then press the ENT key to confirm the selection.

7.If “MR” is selected, the cursor moves to “Range Switch.”

Teledyne Analytical Instruments

35

Page 51
Image 51
Teledyne 7500ZA operating instructions Ndir Gas Analyzer Settings and Calibration, Setting the Range Switch Mode