Teledyne 7500ZA Ndir Gas Analyzer Maintenance Installation Conditions, Sample gas conditions

Models: 7500ZA

1 126
Download 126 pages 9.92 Kb
Page 93
Image 93

NDIR Gas Analyzer

Maintenance

 

 

 

Installation Conditions

1.Ambient temperature: -5°C to +40°C

2.Humidity: 90%RH or less

3.Corrosive gases: None

4.No radiated heat, direct sunlight or rain/wind

5.Dust: No more than local environmental standards permit

6.Vibration: None

Sample gas conditions

1.Temperature: +60°C to +800°C

2.Pressure: -3 to +3 kPa

3.Moisture content: 30% or less

4.Dust: 0.1 g/Nm3 or less

5.Components: 0 to 500 ppm NOX, 0 to 500 ppm SO2, 0 to 2000 ppm CO, 5% to 15% CO2, 0% to 21%O2, 0 to 100 ppm HCl, residue N2

Please consult TAI Customer Service regarding gas analyzer maintenance service requirements.