NDIR Gas Analyzer

Specifications

 

 

 

Electromagnetic Compatibility ;

EN61010-1 : 2010 Safety requirements for electrical

equipment for measurement, control and

laboratory use.

“Installation Category II”

“Pollution Degree 2”

EN61326-1 : 2006

 

Electrical equipment for measurement,

 

control and laboratory use — EMC

 

requirements.

9.1.6 Requirements for Sample Gas

Flow rate:

0.5L / min ±0.2L / min

Temperature:

0 to 50˚C

Pressure:

10 kPa or less (Gas outlet side should be at

 

atmospheric pressure).

Dust:

100 µg/Nm3 or less in particle size of 0.3

 

µm or less.

Mist:

Not allowed

Moisture:

For sample gases NO, SO2, CO(0-200

 

ppm range): less than 2°C saturation point.

 

For most other sample gases: less than

 

standard room temperature saturation

 

point.

Corrosive component:

1 ppm or less

Standard gas for calibration:

 

1)Infrared-ray measurable component, standard O2 Zero gas: Dry N2

Span gas ; Each sample gas having concentration 90 to 100% of its measuring range recommended).

For external zirconia O2 sensor use the same calibration gas line:

Teledyne Analytical Instruments

99

Page 115
Image 115
Teledyne 7500ZA operating instructions Requirements for Sample Gas