NDIR Gas Analyzer | Specifications | |
|
|
|
Electromagnetic Compatibility ;
equipment for measurement, control and
laboratory use.
“Installation Category II”
“Pollution Degree 2”
| Electrical equipment for measurement, |
| control and laboratory use — EMC |
| requirements. |
9.1.6 Requirements for Sample Gas | |
Flow rate: | 0.5L / min ±0.2L / min |
Temperature: | 0 to 50˚C |
Pressure: | 10 kPa or less (Gas outlet side should be at |
| atmospheric pressure). |
Dust: | 100 µg/Nm3 or less in particle size of 0.3 |
| µm or less. |
Mist: | Not allowed |
Moisture: | For sample gases NO, SO2, |
| ppm range): less than 2°C saturation point. |
| For most other sample gases: less than |
| standard room temperature saturation |
| point. |
Corrosive component: | 1 ppm or less |
Standard gas for calibration: |
|
1)
Span gas ; Each sample gas having concentration 90 to 100% of its measuring range recommended).
For external zirconia O2 sensor use the same calibration gas line:
Teledyne Analytical Instruments | 99 |