Settings and Calibration

Model 7500ZA

 

 

 

Cycle : The period between the start time of one calibration and

the next (unit: hour/day)

Flow Time : The time required for purging with calibration gas.

Time required for replacement of sample gas after the calibration is completed (Set by calibration gas. See the next page.)

ON/OFF: Turns the auto calibration feature ON or OFF.

To close the Auto Cal Settings screen or to abort the entry midway, press the ESC key. The previous screen will return.

Gas flow time setting

1.With the cursor on “Flow Time”, press the ENT key to bring up the Flow Time screen.

2.Move the cursor to the gas you want to change by pressing the UP or DOWN key, and then press ENT.

3.The highlighted value can be changed by pressing UP or DOWN key, and then using the SIDE key move the cursor to the right to alter the next digit.

4.After changing the value, press the ENT key to save the setting.

5.Press the ENT key again to

return to the Auto Cal Setting screen.

Note: Only the channels used are displayed on this screen. The Ex. time sets the output signal hold extension time that follows the completion of calibration. It is valid only when the hold setting is set to “ON.” The Ex. time set here is also the hold extension time used in a manual calibration.