NDIR Gas Analyzer | Specifications | |
|
|
|
9.1.2Standard Functions
Output Signal Hold: Output signals are held unchanged during manual and auto calibrations when HOLD function is ON.
The values held are those just before start of calibration mode or setting value. User selectable.
Indication of instantaneous values will not be held.
Range Switching: Range switch function is available in manual, auto, and remote modes. Only preset switch method is effective.
Manual: Allows range to switch by key operation.
Auto: Automatically switched from first range to second range when the measured value exceeds 90%FS of first range. Automatically switched from second range to first range when the measured value drops to 80% or less first range.
Remote (option): Voltage contact input allows the range to switch via an external signal when a remote range switch input is received. When the contact input terminals for each component are input voltage, the first range is selected, and it is switched to the second range when the terminals are open.
*The switch range values are settable between the first range and second range values (low/high range values).