Settings and Calibration

Model 7500ZA

 

 

 

4.On the parameter hold screen that appears, move the cursor to the Ch (component) you want to hold using the UP or DOWN key and then press ENT.

5.A highlighted value indicates that it can be changed. Adjust the highlighted value with the UP or DOWN key and then move the cursor to the right to change the next digit using the SIDE key.

6.After the value is changed, press the ENT key.

Note: The setting is expressed as 1/1 full scale range for both respective ranges.

When 0 to 1000 ppm is selected as the range, and 10% FS is selected as hold setting, the output equivalent to 100 ppm is held irrespective of the measured value at that time.

7.Press the ESC key to return to the parameter setting screen.

Teledyne Analytical Instruments

64