Care and cleaning
Braun ThermoScan probe covers are intended for single use only! To ensure accurate readings, always use Braun ThermoScan disposable probe covers with the Pro 3000 thermometer. Damaged, perforated, or soiled probe covers can lead to inaccurate temperature readings. Additionally, the accuracy of your temperature readings can be affected by damage to the probe window, or the presence of dirt and ear wax on the probe window.
Probe Window
To assure a high level of accuracy, it is very important to keep this window clean, dry and undamaged. Fingerprints, earwax, dust and other soiling compounds reduce transparency of the window and result in lower temperature readings. If the thermometer is accidentally placed in the ear without a probe cover installed, the window must be cleaned immediately. Additionally, the window and/or probe tip area should never be modified, changed, or adjusted. These changes will affect calibration and accuracy of the thermometer.
To clean the window, gently wipe its surface with a cotton swab slightly moistened with alcohol and immediately wipe dry with a clean cotton swab. After cleaning, allow at least 5 minutes drying time before taking temperatures. Make sure probe window is clean and dry. Avoid touching the window except when cleaning is required. If the window is damaged, the thermometer must be returned to Welch Allyn for service.
Note: Do not use any chemical other than alcohol to clean the probe window.
Thermometer
Use a soft, cloth slightly moistened with alcohol to clean the thermometer display and exterior. Do not use abrasive cleaners. Never submerge the thermometer in water or any other liquid.
Store thermometer and probe covers in a dry location free from dust and contamination and away from direct sunlight. The ambient temperature at the storage location should remain fairly constant and within the range of 10°C - 40°C (50°F to 104°F).