B - Performance Test and Calibration

540INPUT "Enter value of high range current shunt in ohms",Rshunt

550DISP "You have entered ";Rshunt;" ohms"

560INPUT "If correct enter 'Y' or press CONT, if wrong enter 'N' or 'n'",Y$

570IF UPC$(Y$)="Y" OR Y$="" THEN 590

580IF UPC$(Y$)="N" OR Y$="n" THEN 520

590CLEAR SCREEN

600!

610OUTPUT @Ld;"CAL:IPR:LEV P1"

620PRINT TABXY(20,20),"Calibrating IMON & IPROG P1 - 30 SECOND WAIT"

630WAIT 30

640CLEAR SCREEN

650!

660INPUT "Measure voltage across current shunt and enter in volts",Ip1

670!

680INPUT "Measure Imon voltage and enter in volts",Vimonp1

690!

700INPUT "Measure Iprog voltage and enter in volts",Viprogp1

710!

720Ip1=Ip1/Rshunt

730!

740OUTPUT @Ld;"CAL:DATA ";Ip1,Vimonp1,Viprogp1

750!

760OUTPUT @Ld;"CAL:IPR:LEV P2"

770!

780PRINT TABXY(20,20),"Calibrating IMON & IPROG P2 - 30 SECOND WAIT"

790WAIT 30

800CLEAR SCREEN

810!

820INPUT "Measure voltage across current shunt and enter in volts",Ip2

830!

840INPUT "Measure Imon voltage and enter in volts",Vimonp2

850!

860INPUT "Measure Iprog voltage and enter in volts",Viprogp2

870!

880Ip2=Ip2/Rshunt

890!

900OUTPUT @Ld;"CAL:DATA ";Ip2,Vimonp2,Viprogp2

910!

920PRINT "Set IPROG power source to 8.5 volts"

930DISP "Press CONT when ready"

940PAUSE

950CLEAR SCREEN

960!

970OUTPUT @Ld;"CAL:IPR:LEV P3"

980PRINT TABXY(20,20),"Calibrating IMON & IPROG P3 - 30 SECOND WAIT"

990WAIT 30

1000 CLEAR SCREEN

1010 !

1020 INPUT "Measure voltage across current shunt and enter in volts",Ip3 1030 !

1040 INPUT "Measure Imon voltage and enter in volts",Vimonp3

1050 !

1060 INPUT "Measure Iprog voltage and enter in volts",Viprogp3

1070 !

1080 Ip3=Ip3/Rshunt

1090 !

1100 OUTPUT @Ld;"CAL:DATA ";Ip3,Vimonp3,Viprogp3

92

Page 89
Image 89
Agilent Technologies n3302a, n3305a, n3304a, n3301a, n3303a, n3306a, n3300a manual Performance Test and Calibration