Test Description | Minimum | Results | Maximum | |
|
| Specification |
| Specification |
| Constant Current Mode Tests |
|
| |
60 Ampere Range Programming and Readback | 0984 A | _______A | 1.016 A | |
Low Current ( 1 A ) |
| |||
Front Panel Display |
| Aout | _______A | Aout + 10.5mA |
High Current ( 60 A ) |
| 59.925 A | _______A | 60.075 A |
Front Panel Display |
| Aout – 40mA | _______A | Aout +40mA |
6 Ampere Range Programming and Readback | 0.9915 | _______A | 1.0085 | |
Low Current ( 1 A ) |
| |||
Front Panel Display |
| Aout – 5.5mA | _______A | Aout + 5.5mA |
High Current ( 6 A ) |
| 5.9865 A | _______A | 6.0135 A |
Front panel Display |
| Aout – 8mA | _______A | Aout + 8mA |
|
|
|
|
|
| Current Regulation Test |
|
| |
5 Amp Regulation at input voltage delta 4V to 60V |
| ______mA | Less then 10 mA | |
|
|
|
|
|
| Voltage Mode Tests |
|
| |
60 Volt Range Programming and Readback | 2.957 V | _______V | 3.043 V | |
Low Voltage ( 3 V ) |
| |||
Front Panel Display |
| Vout – 21mV | _______V | Vout + 21mV |
High Voltage ( 60 V ) |
| 59.932 V | _______V | 60.068 V |
Front Panel Display |
| Vout – 140mV | _______V | Vout + 140mV |
6 Volt Range Programming and Readback | 2.9905 V | _______V | 3.0095 V | |
Low Voltage ( 3 V ) |
| |||
Front Panel Display |
| Vout – 4.5mV | _______V | Vout + 4.5mV |
High Voltage ( 6 V ) |
| 5.991 V | _______V | 6.009 V |
Front Panel Display |
| Vout – 6mV | _______V | Vout + 6mV |
|
|
|
|
|
| Voltage Mode | Regulation Test |
|
|
5V regulation with input current delta 1A to 60A |
| ______mV | Less then 10mV | |
|
|
|
|
|
| Resistance | Mode Tests |
|
|
Resistance Range 1 |
|
| ______ Ω | 2.020 Ω |
High Resistance Point | 2 Ω | 1.980 Ω | ||
Low Resistance Point | 0.50 Ω | 0.486 Ω | ______ Ω | 0.514 Ω |
Resistance Range 2 |
| 1.733 Ω | ______Ω | 1.867 Ω |
Low Resistance Point | 1.8 Ω | |||
High Resistance Point | 20 Ω | 19.360 Ω | ______ Ω | 20.640 Ω |
Resistance Range 3 |
| 16.800 Ω | ______Ω | 19.20 Ω |
Low Resistance Point | 18 Ω | |||
High Resistance Point | 200 Ω | 159.88 Ω | ______Ω | 240.12 Ω |
Resistance Range 4 |
| 144 Ω | ______Ω | 298.8 Ω |
Low Resistance Point | 180 Ω | |||
High Resistance Point | 2000 Ω | 1000 Ω | ______Ω | 42,000 Ω |
|
|
|
|
|
| Transient Generator Tests |
|
| |
Frequency Range Tests (1000 Hz test) | 995 Hz. | _____ Hz | 1005 Hz | |
Duty Cycle Tests (50 % Duty Cycle Test @ 100 Hertz | 49% | ______ % | 51% | |
Pulse Width Test (100∝S wide; 10 A pulse) | 99∝s | ______s | 101∝s | |
|
|
|
|
|