Performance Test and Calibration - B

Agilent N3304A Verification Test Record

Test Description

Minimum

Results

Maximum

 

 

Specification

 

Specification

 

Constant Current Mode Tests

 

 

60 Ampere Range Programming and Readback

0984 A

_______A

1.016 A

Low Current ( 1 A )

 

Front Panel Display

 

Aout –10.5mA

_______A

Aout + 10.5mA

High Current ( 60 A )

 

59.925 A

_______A

60.075 A

Front Panel Display

 

Aout – 40mA

_______A

Aout +40mA

6 Ampere Range Programming and Readback

0.9915

_______A

1.0085

Low Current ( 1 A )

 

Front Panel Display

 

Aout – 5.5mA

_______A

Aout + 5.5mA

High Current ( 6 A )

 

5.9865 A

_______A

6.0135 A

Front panel Display

 

Aout – 8mA

_______A

Aout + 8mA

 

 

 

 

 

 

Current Regulation Test

 

 

5 Amp Regulation at input voltage delta 4V to 60V

 

______mA

Less then 10 mA

 

 

 

 

 

 

Voltage Mode Tests

 

 

60 Volt Range Programming and Readback

2.957 V

_______V

3.043 V

Low Voltage ( 3 V )

 

Front Panel Display

 

Vout – 21mV

_______V

Vout + 21mV

High Voltage ( 60 V )

 

59.932 V

_______V

60.068 V

Front Panel Display

 

Vout – 140mV

_______V

Vout + 140mV

6 Volt Range Programming and Readback

2.9905 V

_______V

3.0095 V

Low Voltage ( 3 V )

 

Front Panel Display

 

Vout – 4.5mV

_______V

Vout + 4.5mV

High Voltage ( 6 V )

 

5.991 V

_______V

6.009 V

Front Panel Display

 

Vout – 6mV

_______V

Vout + 6mV

 

 

 

 

 

 

Voltage Mode

Regulation Test

 

 

5V regulation with input current delta 1A to 60A

 

______mV

Less then 10mV

 

 

 

 

 

 

Resistance

Mode Tests

 

 

Resistance Range 1

 

 

______ Ω

2.020 Ω

High Resistance Point

2 Ω

1.980 Ω

Low Resistance Point

0.50 Ω

0.486 Ω

______ Ω

0.514 Ω

Resistance Range 2

 

1.733 Ω

______Ω

1.867 Ω

Low Resistance Point

1.8 Ω

High Resistance Point

20 Ω

19.360 Ω

______ Ω

20.640 Ω

Resistance Range 3

 

16.800 Ω

______Ω

19.20 Ω

Low Resistance Point

18 Ω

High Resistance Point

200 Ω

159.88 Ω

______Ω

240.12 Ω

Resistance Range 4

 

144 Ω

______Ω

298.8 Ω

Low Resistance Point

180 Ω

High Resistance Point

2000 Ω

1000 Ω

______Ω

42,000 Ω

 

 

 

 

 

 

Transient Generator Tests

 

 

Frequency Range Tests (1000 Hz test)

995 Hz.

_____ Hz

1005 Hz

Duty Cycle Tests (50 % Duty Cycle Test @ 100 Hertz

49%

______ %

51%

Pulse Width Test (100∝S wide; 10 A pulse)

99∝s

______s

101∝s

 

 

 

 

 

85

Page 82
Image 82
Agilent Technologies n3302a, n3305a, n3304a, n3301a, n3303a, n3306a, n3300a manual Agilent N3304A Verification Test Record