Agilent Technologies n3303a, n3305a, n3304a, n3301a, n3306a manual Calibration, Parameters Calibrated

Models: n3303a n3304a n3300a n3302a n3306a n3301a n3305a

1 102
Download 102 pages 27.56 Kb
Page 87
Image 87

B - Performance Test and Calibration

Calibration

Parameters Calibrated

The following parameters may be calibrated.

External Current Monitor ( Imon )

External Current Programming ( Iprog )

Input Current

Input Voltage

Input Resistance

You do not have to do a complete calibration each time. If appropriate, you may calibrate only the current, voltage or resistance and proceed to “Saving Calibration Constants”. However, Imon and Iprog must be calibrated before Input Current is calibrated and both Imon and Input Current must be calibrated before Input Resistance can be calibrated.

The following section lists calibration programs to re-calibrate ( adjust ) those specifications which are not within specification. The programs listed are for use with either an HP Series 200/300 computer using HP Basic ( RMB ) or a PC equipped with an IEEE-488.2 interface card an HP Basic for Windows.

The variables to be used in the programs are listed in Table B-2 Variables. The connections for the test equipment, current shunts and DVMs are shown in Figures B-1 for IMON ZERO calibration, B-2 for IMON, IPROG and CURRENT calibration, B-3 for VOLTAGE calibration and B-4 for RESISTANCE calibration. A listing of all programming commands and their explanations can be found in the Programming Guide, part number 5964-8198.

Table B-2 Variables and Power Source Settings

VARIABLE

N3302A

N3303A

N3304A

N3305A

N3306A

CURRENT MODE VOLTAGE / CURRENT SETTINGS

 

 

 

Power source voltage setting

5 V

20 V

5 V

8 V

5 V

Power source current setting

33 A

12 A

61 A

61 A

130 A

VOLTAGE MODE VOLTAGE /

CURRENTSETTINGS

 

 

 

 

Power source voltage setting

61V

246 V

61V

152V

61V

Power source current setting

2 A

0.6 A

3 A

2 A

3 A

RESISTANCE MODE VARIABLE and VOLTAGE / CURRENT SETTINGS

 

 

Resistance Range 1

 

 

 

 

 

Power source voltage setting for P1

24V

80V

12V

28V

12V

Power source voltage setting for P2

6V

15V

6V

4.5V

6V

Power source current setting

20A

7A

40A

15A

80A

Resistance Range 2

 

 

 

 

 

< resistance range 2 variable >

40 Ω

480 Ω

20 Ω

50 Ω

10 Ω

Power source voltage setting for P1

22V

50V

20V

45V

20V

Power source voltage setting for P2

6V

30V

12V

9V

12V

Power source current setting

3.5A

3.5A

15A

3.5A

15A

Resistance Range 3

 

 

 

 

 

< resistance range 3 variable >

400 Ω

2400 Ω

200 Ω

500 Ω

100 Ω

Power source voltage setting for P1

30V

150V

60V

45V

60V

Power source voltage setting for P2

15V

75V

60V

15V

60V

Power source current setting

3.5A

3.5A

3.5A

3.5

7A

Resistance Range 4

 

 

 

 

 

< resistance range 4 variable >

2000 Ω

12000 Ω

2000 Ω

2500 Ω

1000 Ω

Power source voltage setting for P1

60V

150V

60V

150V

18V

Power source voltage setting for P2

30V

150V

60V

115V

18V

Power source current setting

3.5A

3.5A

3.5A

3.5A

13.5A

90

Page 87
Image 87
Agilent Technologies n3303a, n3305a manual Calibration, Parameters Calibrated, Table B-2 Variables and Power Source Settings