B - Performance Test and Calibration

Agilent N3305A Verification Test Record

Test Description

 

Minimum

Results

Maximum

 

 

Specification

 

Specification

Constant Current Mode Tests

 

 

60 Ampere Range Programming and Readback

0.984 A

_______A

1.016 A

Low Current ( 1A )

 

Front Panel Display

 

Aout – 10.5mA

_______A

Aout + 10.5mA

High Current ( 60 A )

 

59.925 A

_______A

60.075 A

Front Panel Display

 

Aout – 40mA

_______A

Aout + 40mA

6 Ampere Range Programming and Readback

 

0.9915 A

_______A

1.0085 A

Low Current ( 1A )

 

Front Panel Display

 

Aout – 5.5mA

_______A

Aout + 5.5mA

High Current ( 6 A )

 

5.9865 A

_______A

6.0135 A

Front panel Display

 

Aout – 8mA

_______A

Aout + 8mA

 

 

 

 

 

 

Current Regulation Test

 

 

3.3 Amp Regulation at input voltage delta 4V to 60V

 

______mA

Less then 10mA

 

 

 

 

 

 

Voltage Mode Tests

 

 

150 Volt Range Programming and Readback

 

2.977 V

_______V

3.023 V

Low Voltage ( 3 V )

 

Front Panel Display

 

Vout – 17.5mV

_______V

Vout + 17.5mV

High Voltage ( 150 V )

 

149.830 V

_______V

150.170 V

Front Panel Display

 

Vout – 9.1mV

_______V

Vout + 9.1mV

15 Volt Range Programming and Readback

 

2.987 V

_______V

3.013 V

Low Voltage ( 3 V )

 

Front Panel Display

 

Vout – 9.5mV

_______V

Vout + 9.5mV

High Voltage ( 15 V )

 

14.975 V

_______V

15.025 V

Front Panel Display

 

Vout – 15.5mV

_______V

Vout + 15.5mV

 

 

 

 

 

 

Voltage Regulation Test

 

 

8.3V regulation with input current delta 1A to 60A

 

______mV

Less then 10mV

 

 

 

 

 

Resistance Mode Tests

Resistance Range 1

 

High Resistance Point

5 Ω

Low Resistance Point

0.125 Ω

Resistance Range 2

 

Low Resistance Point

4.5 Ω

High Resistance Point

50 Ω

Resistance Range 3

 

Low Resistance Point

45 Ω

High Resistance Point

500 Ω

Resistance Range 4

 

Low Resistance Point

450 Ω

High Resistance Point

2500 Ω

4.936 Ω

0.1005 Ω

4.352 Ω

48.42

40.50

350 Ω

292

1000 Ω

______ Ω

______ Ω

______Ω

______Ω

______Ω

______Ω

______Ω

______Ω

5.064 Ω

0.1495 Ω

4.647 Ω

51.58

56.25

1250 Ω

1575 Ω

252 k Ω

Transient Generator Tests

Frequency Range Tests (1000 Hz test)

995 Hz.

_____ Hz

1005 Hz

Duty Cycle Tests (50 % Duty Cycle Test @ 100 Hertz

49%

______ %

51%

Pulse Width Test (100∝S wide; 10 A pulse)

99∝s

______s

101∝s

 

 

 

 

86

Page 83
Image 83
Agilent Technologies n3300a, n3305a, n3304a, n3301a, n3303a, n3306a, n3302a manual Agilent N3305A Verification Test Record