Agilent Technologies n3305a, n3304a, n3301a, n3303a, n3306a Agilent N3306A Verification Test Record

Models: n3303a n3304a n3300a n3302a n3306a n3301a n3305a

1 102
Download 102 pages 27.56 Kb
Page 84
Image 84

Performance Test and Calibration - B

Agilent N3306A Verification Test Record

Test Description

 

Minimum

Results

Maximum

 

 

Specification

 

Specification

 

Constant Current Mode Tests

 

 

120 Ampere Range Programming and Readback

0.9615 A

_______A

1.0385 A

Low Current ( 1A )

 

Front Panel Display

 

Aout – 20.5mA

_______A

Aout + 20.5mA

High Current ( 120 A )

 

119.8425 A

_______A

120.1575 A

Front Panel Display

 

Aout – 80mA

_______A

Aout + 80mA

12 Ampere Range Programming and Readback

0.9840A

_______A

1.0160 A

Low Current ( 1A )

 

Front Panel Display

 

Aout – 10.5mA

_______A

Aout + 10.5mA

High Current ( 12 A )

 

11.973 A

_______A

12.027 A

Front panel Display

 

Aout – 16mA

_______A

Aout + 16mA

 

 

 

 

 

 

Current Regulation Test

 

 

10 Amp Regulation at input voltage delta 4V to 60V

 

______mA

Less then 10mA

 

 

 

 

 

 

Voltage Mode Tests

 

 

60 Volt Range Programming and Readback

 

2.989 V

_______V

3.011 V

Low Voltage ( 3 V )

 

Front Panel Display

 

Vout – 21mV

_______V

Vout + 21mV

High Voltage ( 60 V )

 

59.932 V

_______V

60.068 V

Front Panel Display

 

Vout – 38mV

_______V

Vout + 38mV

6 Volt Range Programming and Readback

 

2.9905 V

_______V

3.0095 V

Low Voltage ( 3 V )

 

Front Panel Display

 

Vout – 4.5m

_______V

Vout + 4.5m

High Voltage ( 6 V )

 

5.991 V

_______V

6.009 V

Front Panel Display

 

Vout – 6mV

_______V

Vout + 6mV

 

 

 

 

 

 

Voltage Regulation Test

 

 

5V regulation with input current delta 1A to 120A

 

______mV

Less then 20mV

 

 

 

 

 

Resistance Mode Tests

Resistance Range 1

 

High Resistance Point

1 Ω

Low Resistance Point

0.100 Ω

Resistance Range 2

 

Low Resistance Point

0.9 Ω

High Resistance Point

10 Ω

Resistance Range 3

 

Low Resistance Point

9 Ω

High Resistance Point

100 Ω

Resistance Range 4

 

Low Resistance Point

90 Ω

High Resistance Point

1000 Ω

0.990 Ω

0.0936 Ω

0.8665 Ω

9.680 Ω

8.400 Ω

79.940 Ω

72

500 Ω

______ Ω

______ Ω

______Ω

______Ω

______Ω

______Ω

______Ω

______ Ω

1.010 Ω

0.1064 Ω

0.9335 Ω

10.32

9.60

120.06 Ω

149.4

21,000 Ω

Transient Generator Tests

Frequency Range Tests (1000 Hz test)

995 Hz.

_____ Hz

1005 Hz

Duty Cycle Tests (50 % Duty Cycle Test @ 100 Hertz

49%

______ %

51%

Pulse Width Test (100∝S wide; 10 A pulse)

99∝s

______s

101∝s

 

 

 

 

87

Page 84
Image 84
Agilent Technologies n3305a, n3304a, n3301a, n3303a, n3306a, n3302a, n3300a manual Agilent N3306A Verification Test Record