B - Performance Test and Calibration

6.

Set power source voltage and current values as listed in following table.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

 

N3303A

 

N3304A

N3305A

N3306A

 

Power Source Voltage Setting

24V

 

96V

 

12V

30V

12V

 

 

Power Source Current Setting

7A

 

3A

 

7A

15A

8A

 

7.

Wait 15 seconds. Calculate and record resistance range 2 high resistance point

 

 

 

 

 

Resistance Range 3 Programming Accuracy

 

 

 

 

 

 

 

1.

Press [ RES ]. Scroll until display reads RES:RANG, for models N3302A, N3304A, N3305A and N3306A press

 

[ 1 ] [ 0 ] [ 0 ] then [ Enter ]. For model N3302A press [ 5 ] [ 0 ]

[ 0 ] then [ Enter ].

 

 

 

2.

Checking resistance range 3, low resistance point. Press [ RES ] [select resistance from following table ] [ Enter ].

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

N3303A

 

N3304A

 

 

N3305A

 

N3306A

 

 

 

 

36 Ω

440 Ω

 

18 Ω

 

 

45 Ω

 

9 Ω

 

 

 

3.

Set power source voltage and current values as listed in following table.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

 

N3303A

 

N3304A

N3305A

N3306A

 

Power Source Voltage Setting

40V

 

96V

 

20V

50V

20V

 

 

Power Source Current Setting

3.5A

 

1A

 

3A

3.5A

3.5A

 

4.

Wait 7 seconds. Calculate and record resistance range 3 low resistance point.

 

 

 

 

 

5.

Checking resistance range 3, high resistance point. Press [ RES ] [select resistance from following table ] [ Enter ].

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

N3303A

 

N3304A

 

 

N3305A

 

N3306A

 

 

 

 

400 Ω

4800 Ω

 

200 Ω

 

 

500 Ω

 

100 Ω

 

 

 

6.

Set power source voltage and current values as listed in following table.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

 

N3303A

 

N3304A

N3305A

N3306A

 

Power Source Voltage Setting

40V

 

96V

 

20V

50V

20V

 

 

Power Source Current Setting

3.5A

 

1A

 

3A

3.5A

3.5A

 

7.

Wait 7 seconds. Calculate and record resistance range 3 high resistance point.

 

 

 

 

 

Resistance Range 4 Programming Accuracy

 

 

 

 

 

 

 

1.

Press [ RES ]. Scroll until display reads RES:RANG, For models N3302A, N3304A, N3305A and N3306A press

 

[ 1 ] [ 0 ] [ 0 ] [ 0 ] then [ Enter ]. For model N3303A press [ 5 ] [ 0 ] [ 0 ] [ 0 ] then [ Enter ].

 

 

 

2.

Checking resistance range 4 low resistance point. Press [ RES ] [select resistance from following table ] [ Enter ].

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

N3303A

 

N3304A

 

 

N3305A

 

N3306A

 

 

 

 

360 Ω

4400 Ω

 

180 Ω

 

 

450 Ω

 

90 Ω

 

 

 

3.

Set power source voltage and current values as listed in following table.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

 

N3303A

 

N3304A

N3305A

N3306A

 

Power Source Voltage Setting

50V

 

96V

 

20V

50V

20V

 

 

Power Source Current Setting

3.5A

 

1A

 

3A

3.5A

3.5A

 

4.

Wait 3 seconds. Calculate and record resistance range 3 low resistance point.

 

 

 

 

 

5.

Checking resistance range 4, high resistance point. Press [ RES ] [select resistance from following table ] [ Enter ].

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

N3303A

 

N3304A

 

 

N3305A

 

N3306A

 

 

 

 

2000 Ω

12,000 Ω

 

2000 Ω

 

 

2500 Ω

 

1000 Ω

 

 

 

80