Agilent Technologies n3305a, n3304a, n3301a, n3303a, n3306a, n3302a Performance Test and Calibration

Models: n3303a n3304a n3300a n3302a n3306a n3301a n3305a

1 102
Download 102 pages 27.56 Kb
Page 91
Image 91

B - Performance Test and Calibration

1690 CLEAR SCREEN

1700 !

1710 OUTPUT @Ld;"INP:STAT ON"

1720 OUTPUT @Ld;"CAL:LEV P1"

1730 !

1740 PRINT TABXY(10,10),"Calibrating low current range P1 - 30 SECOND WAIT" 1750 WAIT 30

1760 CLEAR SCREEN

1770 !

1780 INPUT "Measure voltage across current shunt and enter in volts",Ip1 1790 !

1800 Ip1=Ip1/Rshunt

1810 !

1820 OUTPUT @Ld;"CAL:DATA ";Ip1

1830 !

1840 OUTPUT @Ld;"CAL:LEV P2"

1850 !

1860 PRINT TABXY(10,10),"calibrating low current range P2 - 30 SECOND WAIT" 1870 WAIT 30

1880 CLEAR SCREEN

1890 !

1900 INPUT "Measure voltage across current shunt and enter in volts",Ip2 1910 !

1920 Ip2=Ip2/Rshunt

1930 !

1940 OUTPUT @Ld;"CAL:DATA ";Ip2

1950 !

1960 OUTPUT @Ld;"CAL:SAVE"

1970 OUTPUT @Ld;"CAL:STAT OFF"

1980 !

1990 PRINT "IMON, IPROG and Current calibration adjustments complete"

2000 PRINT "Verify Current Adjustments"

2010 !

2020 END

94

Page 91
Image 91
Agilent Technologies n3305a, n3304a, n3301a, n3303a, n3306a, n3302a, n3300a manual Performance Test and Calibration