B - Performance Test and Calibration

540 !

550INPUT "Measure voltage at +/- sense terminals and enter value in volts",Vp1

560!

570OUTPUT @Ld;"CAL:DATA";Vp1

580!

590OUTPUT @Ld;"CAL:LEV P2"

600WAIT 1

610!

620INPUT "Measure voltage at +/- sense terminals and enter value in volts",Vp2

630!

640OUTPUT @Ld;"CAL:DATA";Vp2

650!

660OUTPUT @Ld;"CAL:SAVE"

670OUTPUT @Ld;"CAL:STAT OFF"

680!

690CLEAR SCREEN

700PRINT "Voltage calibration adjustments complete"

710PRINT "Verify Voltage Adjustments"

720!

730END

96

Page 93
Image 93
Agilent Technologies n3301a, n3305a, n3304a, n3303a, n3306a, n3302a, n3300a manual Performance Test and Calibration