Performance Test and Calibration - B

CR Mode Tests

These tests verify that the module operates in CR mode and the resistance programming is within specification. The programmed resistance values are checked by recording the voltage across the current monitor resistor and the input voltage (voltage at the modules + and – sense terminals ), then calculating the resistance as follows;

Load Resistance = Input Voltage / ( voltage across current monitor / monitor resistor value )

Resistance Range 1 Programming Accuracy

1.

Connect Electronic Load, Power source and DVM’s as shown in figure B-4. Be sure + sense and – sense are

 

connected to the +/- input terminals and the remote/local button is in the remote position.

 

2.

Turn on the Electronic Load and press [ Recall ], scroll till display reads *RST and press [ Enter ].

 

3.

Press [ Func ]. Scroll until display reads FUNC RES then press [ Enter ].

 

 

 

4.

Press [ RES ]. Scroll until display reads RES:RANG, press [ 1 ] then [ Enter ].

 

 

 

5.

Turn on the power source. See power source voltage and current values as listed in following table.

 

 

 

 

 

 

 

 

 

 

N3302A

N3303A

N3304A

N3305A

 

N3306A

 

Power Source Voltage Setting

24V

48V

12V

30V

 

15V

 

Power Source Current Setting

7A

7A

10A

10A

 

20A

6.

Wait 30 seconds. Checking resistance range 1, high resistance point. Calculate and record resistance on test card.

 

Input resistance = Input Voltage / ( DVM voltage reading from current shunt/ current shunt resistance).

7.

Checking resistance range 1, low resistance point. Reset power source voltage to following values.

 

 

 

 

 

 

 

 

 

 

N3302A

N3303A

N3304A

N3305A

 

N3306A

 

Power Source Voltage Setting

4V

6V

10V

5V

 

6V

 

Power Source Current Setting

25A

10A

30A

50A

 

70A

8.

Press [ RES ] [ select resistance from following table ] [ Enter ].

 

 

 

 

 

 

 

 

 

 

 

N3302A

N3303A

N3304A

N3305A

N3306A

 

0.2Ω

1.2 Ω

0.5 Ω

0.125 Ω

0.100 Ω

 

9.

Wait 30 seconds. Calculate and record resistance range 1, low resistance point.

 

 

Resistance Range 2 Programming Accuracy

 

 

 

1.

Press [ RES ]. Scroll until display reads RES:RANG, For models N3302A, N3304A, N3305A and N3306A press

 

[ 9 ] then [ Enter ]. For model N3303A Press [ 5 ] [ 0 ] then [ Enter ].

 

 

2.

Checking resistance range 2, low resistance point. Press [ RES ] [select resistance from following table ] [ Enter ].

 

 

 

 

 

 

 

 

N3302A

N3303A

N3304A

N3305A

N3306A

 

3.6 Ω

44 Ω

1.8 Ω

4.5 Ω

0.9 Ω

 

3. Set power source voltage and current values as listed in following table.

 

 

 

 

 

 

 

 

 

 

 

N3302A

N3303A

N3304A

N3305A

N3306A

 

Power Source Voltage Setting

20V

96V

6V

30V

6V

 

Power Source Current Setting

7A

3A

7A

15A

8A

4.

Wait 15 seconds. Calculate and record resistance range 2 low resistance point.

 

 

5.

Checking resistance range 2, high resistance point. Press [ RES ] [select resistance from following table ] [ Enter ].

 

 

 

 

 

 

 

 

N3302A

N3303A

N3304A

N3305A

N3306A

 

40 Ω

480 Ω

20 Ω

50 Ω

10 Ω

 

79

Page 76
Image 76
Agilent Technologies n3300a, n3305a, n3304a, n3301a, n3303a, n3306a CR Mode Tests, Resistance Range 1 Programming Accuracy