Agilent Technologies n3306a, n3305a, n3304a, n3301a IMON, Iprog and Current Calibration Program

Models: n3303a n3304a n3300a n3302a n3306a n3301a n3305a

1 102
Download 102 pages 27.56 Kb
Page 88
Image 88

Performance Test and Calibration - B

IMON, IPROG and CURRENT Calibration Program

10 ! This program will calibrate Imon, Iprog and Current

20! for load modules N3302A, N3303A, N3304A, N3305A and N3306A.

30!

40! last update 9/6/00

50!

60! Edit the interface address on line 100 if different then 705

70!

80ASSIGN @Ld TO 705

90!

100OUTPUT @Ld;"*RST"

110OUTPUT @Ld;"CAL:STAT ON"

120OUTPUT @Ld;"FUNC CURR"

130OUTPUT @Ld;"CURR:RANG MAX"

140OUTPUT @Ld;"SENS:CURR:RANG MAX"

150OUTPUT @Ld;"INP:STAT ON"

160!

170

!

*******

Calibrating Imon Zero *******

180

!

 

 

190

PRINT TABXY(10,8),"See figure B-1 for Imon calibration"

200PRINT TABXY(10,10),"Module Input Terminals must be open - disconnect power source"

210PRINT TABXY(10,12),"Connect DVM to IMON terminals"

220DISP "Press CONT when ready to calibrate Imon"

230PAUSE

240CLEAR SCREEN

250!

260OUTPUT @Ld;"CAL:IMON:LEV P1"

270PRINT TABXY(20,20);"Calibrating IMON P1 - 5 SECOND WAIT"

280WAIT 5

290CLEAR SCREEN

300INPUT "Measure IMON and enter value in volts - OBSERVE VOLTAGE POLARITY",Vimonp1

310!

320OUTPUT @Ld;"CAL:DATA ";Vimonp1

330!

340OUTPUT @Ld;"CAL:IMON:LEV P2"

350PRINT TABXY(20,20);"Calibrating IMON P2 - 5 SECOND WAIT"

360WAIT 5

370CLEAR SCREEN

380INPUT "Measure IMON and enter value in volts - OBSERVE VOLTAGE POLARITY",Vimonp2

390!

400OUTPUT @Ld;"CAL:DATA ";Vimonp2

410!

420

!

*******

Calibrating Imon and Iprog gain *******

430

!

 

 

440PRINT TABXY(10,8),"See figure B-2 for Imon and Iprog calibration"

450PRINT TABXY(10,10),"Connect module input terminals to power source"

460PRINT TABXY(10,12),"See Table B-2 Current Mode Voltage /Current Settings"

470PRINT TABXY(10,14),"Set power source voltage and current for model to be calibrated"

480PRINT TABXY(10,16),"Set IPROG power source to 1 volt"

490!

500DISP "Press CONT when ready to calibrate Imon and Iprog"

510PAUSE

520CLEAR SCREEN

530!