Performance Test and Calibration - B

Agilent N3302A Verification Test Record

Test Description

Minimum

Results

Maximum

 

 

Specification

 

Specification

 

Constant Current Mode Tests

 

 

30 Ampere Range Programming and Readback

0.989 A

_______A

1.011 A

Low Current ( 1 A )

 

Front Panel Display

 

Aout – 6.5mA

_______A

Aout + 6.5mA

High Current ( 30 A )

 

29.960 A

_______A

30.040 A

Front Panel Display

 

Aout – 21mA

_______A

Aout + 21mA

3 Ampere Range Programming and Readback

0.994 A

_______A

1.006 A

Low Current ( 1 A )

 

Front Panel Display

 

Aout – 3.5mA

_______A

Aout + 3.5mA

High Current ( 3 A )

 

2.992 A

_______A

3.008 A

Front panel Display

 

Aout – 4.5mA

_______A

Aout + 4.5mA

 

Constant Current

Regulation Test

 

 

2.5 Amp Regulation at input voltage delta 4V to 60V

 

______ mA

Less then 10mA

 

 

 

 

 

 

Voltage Mode Tests

 

 

60 Volt Range Programming and Readback

2.989 V

_______V

3.011 V

Low Voltage ( 3 V )

 

Front Panel Display

 

Vout – 9.5mV

_______V

Vout + 9.5mV

High Voltage ( 60 V )

 

59.932 V

_______V

60.068 V

Front Panel Display

 

Vout – 3.8mV

_______V

Vout + 3.8mV

6 Volt Range Programming and Readback

2.9905 V

_______V

3.0095 V

Low Voltage ( 3 V )

 

Front Panel Display

 

Vout – 4.5mV

_______V

Vout + 4.5mV

High Voltage ( 6 V )

 

5.991 V

_______V

6.009 V

Front Panel Display

 

Vout – 6.0mV

_______V

Vout + 6.0mV

 

 

 

 

 

 

Voltage Mode

Regulation Test

 

 

5V regulation with input current delta 1A to 30A

 

______mV

Less then 5mV

 

 

 

 

 

 

Resistance

Mode Tests

 

 

Resistance Range 1

 

3.944 Ω

______ Ω

4.056 Ω

High Resistance Point

4 Ω

Low Resistance Point

0.1 Ω

0.0756 Ω

______ Ω

0.1244 Ω

Resistance Range 2

 

 

 

 

Low Resistance Point

3.6 Ω

3.466 Ω

______Ω

3.734 Ω

High Resistance Point

40 Ω

38.520 Ω

______Ω

41.480 Ω

Resistance Range 3

 

 

 

 

Low Resistance Point

36 Ω

32.880 Ω

______Ω

39.120 Ω

High Resistance Point

400 Ω

271.76 Ω

______ Ω

528.24 Ω

Resistance Range 4

 

 

 

 

Low Resistance Point

360 Ω

270 Ω

______Ω

720 Ω

High Resistance Point

2000 Ω

1200 Ω

______Ω

10,000 Ω

 

 

 

 

 

 

Transient Generator Tests

 

 

Frequency Range Tests (1000 Hz test)

995 Hz.

_____ Hz

1005 Hz

Duty Cycle Tests (50 % Duty Cycle Test @ 100 Hertz

49%

______ %

51%

Pulse Width Test (100∝S wide; 10 A pulse)

99∝s

______s

101∝s

 

 

 

 

 

83

Page 80
Image 80
Agilent Technologies n3303a, n3305a, n3304a, n3301a, n3306a, n3302a, n3300a manual Agilent N3302A Verification Test Record