Front Panel Operation - 5

Scrolling Keys

QR

QR

Step Step

Q

Channel

Scrolling keys let you move through the commands in the presently selected function menu. Press ▼ to bring up the next command in the list. Press ▲ to go back to the previous command in the list. Function menus are circular; you can return to the starting position by continuously pressing either key

These keys let you scroll through points in a list function. When the display indicates EOL, the end of the list has been reached.

Press this key to scroll through all channels in the mainframe

Metering Keys

Metering keys control the metering functions of the electronic load. As set from the factory, all front panel measurements are calculated from a total of 1000 readings taken at a 10 microsecond sampling rate. Therefore, the factory default acquisition time for a single front panel measurement is about 10 milliseconds. Refer to “Making Front Panel Measurements” for more information about changing the front panel sampling rate and the number of measurement points.

Meter

Takes the front panel back to the Metering mode from any other mode.

Also accesses the meter menu. This menu lets you measure various parameters of the selected input.

Display

Measurement

<chan> <voltage> <current>

Displays the voltage and current of the selected channel

<chan> <value> V MAX

Displays the maximum voltage

<chan> <value> V MIN

Displays the minimum voltage

<chan> <value> V RMS

Displays the rms voltage

<chan> <value> A MAX

Displays the maximum current

<chan> <value> A MIN

Displays the minimum current

<chan> <value> A RMS

Displays the rms voltage

<chan> <value> WATTS

Displays the input power

<chan> <value> W MAX

Displays the maximum power

<chan> <value> W MIN

Displays the minimum power

Shift

Sense

Specifies the measurement functions and ranges.

 

 

Display

Command Function

 

 

S:PNT

Defines the number of data points in the measurement

 

 

S:TIN

Sets the digitizer sample spacing

 

 

S:WIN

Sets the measurement window function

 

 

S:OFF

Defines the data offset in the measurement

 

 

S:C:RNG

Selects the current measurement range

 

 

S:V:RNG

Selects the voltage measurement range

Notes:Use

Use

R

RInput

and and

Q

QInput

to scroll through the menu commands

to scroll through the command parameters

59