Performance Test and Calibration - B

VOLTAGE Calibration Program

10! This program calibrates the voltage mode for load modules

20! N3302A, N3303A, N3304A, N3305A and N3306A.

30!

40! last update 9/6/00

50!

60! Edit the interface address on line 100 if different then 705

70!

80ASSIGN @Ld TO 705

90!

100OUTPUT @Ld;"*RST"

110OUTPUT @Ld;"CAL:STAT ON"

120OUTPUT @Ld;"FUNC VOLT"

130OUTPUT @Ld;"VOLT:RANG MAX"

140OUTPUT @Ld;"SENS:VOLT:RANG MAX"

150OUTPUT @Ld;"INP:STAT ON"

160!

170

!

*******

Calibrating high volt range

*******

180

!

 

 

 

190PRINT TABXY(10,8),"See figure B-3 for voltage calibration"

200PRINT TABXY(10,10),"Connect power leads to input terminals"

210PRINT TABXY(10,12),"Connect + sense to + input terminal"

220PRINT TABXY(10,14),"Connect - sense to - input terminal"

230PRINT TABXY(10,16),"Connect DVM to +/- sense terminals"

240PRINT TABXY(10,18),"See Table B-2 Voltage Mode Voltage / Current Settings"

250PRINT TABXY(10,20),"Set power source to voltage and current for model to be calibrated"

260PRINT TABXY(10,24),"BE SURE REMOTE BUTTON ON REAR PANEL IS IN REMOTE POSITION"

270DISP "Press CONT when ready to calibrate voltage "

280PAUSE

290CLEAR SCREEN

300!

310PRINT TABXY(9,15);"CALIBRATING HIGH VOLTAGE RANGE"

320OUTPUT @Ld;"CAL:LEV P1"

330WAIT 1

340INPUT "Measure voltage at +/- sense terminals and enter value in volts",Vp1

350!

360OUTPUT @Ld;"CAL:DATA";Vp1

370!

380OUTPUT @Ld;"CAL:LEV P2"

390WAIT 1

400INPUT "Measure voltage at +/- sense terminals and enter value in volts",Vp2

410!

420OUTPUT @Ld;"CAL:DATA";Vp2

430!

440

!

*******

Calibrating low voltage range *******

450

!

 

 

460

OUTPUT @Ld;"VOLT:RANG MIN"

470OUTPUT @Ld;"SENS:VOLT:RANG MIN"

480!

490CLEAR SCREEN

500PRINT TABXY(9,15);"CALIBRATING LOW VOLTAGE RANGE"

510!

520OUTPUT @Ld;"CAL:LEV P1"

530WAIT 1

95

Page 92