B - Performance Test and Calibration

4. Turn on the power source. Set the power source voltage and current to the values listed in the following table.

 

 

 

 

 

 

 

 

 

N3302A

N3303A

N3304A

N3305A

N3306A

 

Power Source Voltage Setting

61V

246V

61V

152V

61V

 

Power Source Current Setting

2A

0.6A

3A

2A

3A

5.

Checking high voltage range, high voltage point. Wait 10 seconds then record actual input voltage and front panel

 

 

voltage reading on test record card under high voltage range, high voltage.

 

 

 

6.

Checking high voltage range, low voltage point. Press [ VOLT ]

[ 3 ] [ Enter ]. Wait 10 seconds, then record actual

 

 

input voltage and front panel readings on test record card under high voltage range, low voltage.

 

7.

Checking low voltage range, low voltage point. Press [ VOLT ]. Scroll until display reads V:RANG Press [ 1 ]

[

 

Enter ]. Press shift key then [ Sense ] scroll till display reads S:V:RNG, press [ 3 ]. Wait 10 seconds then record actual

 

input voltage and front panel readings on test record card under low voltage range, low voltage.

 

8.

Checking low voltage range, high voltage point. Press [ VOLT ] [ select voltage from following table ] [ Enter ].

 

 

 

 

 

 

 

 

 

 

 

N3302A

N3303A

N3304A

 

N3305A

N3306A

 

 

6V

24V

6V

 

15V

6V

 

 

9.

Wait 10 seconds then record actual input voltage and front panel readings on test record card under low voltage range,

 

low voltage.

 

 

 

 

 

 

 

CV Mode Regulation Test

This test verifies the Input Voltage remains within specification when the input current is changed from a low current to rated current.

1.

Connect Electronic Load, power source and DVM as shown in figure B-3. Be sure + sense and – sense are connected

 

to the +/- input terminals and the remote/local button is in the remote position.

 

 

 

2.

Turn on the Electronic Load and press [ Recall ]. Scroll until display reads *RST and press [ Enter ].

3.

Press [ Func ]. Scroll until display reads FUNC VOLT, then press [ Enter ].

 

 

 

 

 

4.

Press [ Volt ] [ select voltage from following table ] [ Enter ].

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

N3303A

 

N3304A

 

N3305A

 

N3306A

 

 

 

 

5V

25V

 

5V

 

8.3V

 

5V

 

 

 

5.

Turn on power source. Set power source voltage and current to values listed in following table.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

 

N3303A

 

N3304A

N3305A

N3306A

 

Power Source Voltage Setting

6V

 

30V

 

6.5V

10V

6.6V

 

 

Power Source Current Setting

1A

 

0.4A

 

1A

1A

1A

 

6.

Wait 10 seconds and record input voltage reading.

 

 

 

 

 

 

 

 

7.

Reset power source current level to value listed in following table.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N3302A

 

N3303A

 

N3304A

N3305A

N3306A

 

Power Source Current Setting

30A

 

10A

 

60A

60A

120A

 

8.

Wait 10 seconds and record input voltage reading.

 

 

 

 

 

 

 

 

9.

Subtract reading in step ‘8’ from reading in step ‘6’. Absolute value of difference should be less then specification.

78